SGC and SGC Plus User’s Manual

UM SGC v3.0

7 of 15

 If the self-test fails, the detector emits five short beeps and flashes before displaying “Test”. Repeat the self-test.

 If the self-test fails three (3) consecutive times the detector will enter a Fail Safe mode. Please contact GCT for a 

replacement detector.

 During normal operation, the battery is continuously monitored. If the battery is low for more than (3) hours the 

detector enters Fail Safe mode.

 If the battery self-test fails (5) consecutive times the LCD will display “EO5”. In case of an EO5 screen discontinue 

use and contact GCT for a replacement detector.

 Along with the battery, the sensor is continuously monitored during normal operation. If a problem with the 

sensor is detected, the screen will display “EO6”. In case of an EO6 screen, please discontinue use and contact GCT 

for a replacement detector.

 If the detector is displaying “bUP”, the detector is either due for a bump test because of a scheduled test or has 

failed a bump test. Please refer to “Bump Test Interval” for more detail.

 If the detector is displaying “EOL” it has reached the end of its operating life. Please discontinue use.

Alarm Types:

LOW ALARM

Audible alarm: One (1) slow beep every second

Visual alarm: One (1) slow flash every second

Vibrating alarm: One (1) slow vibration every second

HIGH ALARM and OVER LIMIT (OL) ALARM

Audible alarm: Two (2) fast beeps every second

Visual alarm: Two (2) fast flashes every second

Vibrating alarm: Two (2) fast vibrations every second

DETECTOR LIFE COUNTDOWN ALERT “EOL”

Once the detector has less than one (1) month of life 

remaining the screen will switch to days remaining. When 

there is less than one (1) day remaining the screen will 

switch to hours remaining.  Once the detector has eight 

(8) hours remaining it will begin to beep, flash, and vibrate 

intermittently. To end the alert push the button down. Once 

the detector has reached the end of its operating life the 

display will show “EOL” (End of Life).

ALARMS

When a particular gas has a Gas Reading (4) at, or above, its alarm thresholds, the Gas Identifier (5) icon will flash and 

the associated Alarm Condition (1) icon will display.

During a calibration or bump test, the Calibration/Test Mode (2) icon will display when it is time to apply gas.

The Battery Charge Level (3) is displayed in a 3 bar battery icon as well as a percentage. The percentage calculation is 

approximate and can be used to provide a rough estimation of the time remaining.

Summary of Contents for Multi Gas Clip

Page 1: ...Single Gas Clip Plus Single Gas Clip User s Manual G A S C L I P T E C H C O M...

Page 2: ...n 5 Turning the Detector On 5 Turning the Detector Off 5 Performing a Self test 5 Alarms 7 Detector Maintenance 8 Bump Test Interval bUP 8 O2 Detector Calibration 8 Cleaning 9 Hibernation SGC Plus Onl...

Page 3: ...ion The CO and H2S versions of both the Single Gas Clip and SGC Plus do not require calibration for the life of the product The detector should be bump tested before each day s use with a known concen...

Page 4: ...ur l talonnage Les versions CO et H2S du Single Gas Clip et du SGC Plus ne n cessitent pas d talonnage pendant toute la dur e de vie du produit Le d tecteur doit tre test avant chaque utilisation quot...

Page 5: ...of 15 DETECTOR COMPONENTS ENTRY DESCRIPTION 1 Audible Alarm Port 2 Display 3 Alarm Bar LEDs 4 Maintenance LEDs 5 Power Button 6 Sensor 7 Infrared IR Communication Window 8 Alligator Clip with Safety...

Page 6: ...play Details The detector utilizes a special high viewing angle LCD that is designed to enhance the screen visibility In the absence of gas it displays life remaining In those cases where gas is prese...

Page 7: ...and the sensor is tested Below we outlined the most common screen configurations however if the detector has been programmed via the GCT IR Link or has been exposed to gas additional screens may appea...

Page 8: ...ctor will be reset Note while the value will be cleared on the display the value will be held in the detector s log See the event log section for further details This value can be cleared on the next...

Page 9: ...life Please discontinue use Alarm Types LOW ALARM Audible alarm One 1 slow beep every second Visual alarm One 1 slow flash every second Vibrating alarm One 1 slow vibration every second HIGH ALARM an...

Page 10: ...button press to skip the bump test If the SGC is bumped while showing GAS then it will record as a bump test in the event log instead of as an exposure If no gas is applied it will return to the norma...

Page 11: ...ve the event log by pressing the save event log before hibernating 4 Click on the Hibernate Button acknowledge the event log message 5 Keep in front of the GCT IR Link until the Hibernate OK message i...

Page 12: ...of logs you want to download a Event Logs Only b Events Partial Data Log last week of data c Events New Data Logs Only All data logs from last time you pulled the logs to current d Full Logs approxim...

Page 13: ...CALCAP Replacement calibration cap for the Single Gas Clip or the Single Gas Clip Plus Single Gas Clip Dock P N SGC DOCK portable chargeable all in one docking station in a durable portable case for a...

Page 14: ...ts Newer events replace older events Warranty Full 2 years SGC or 3 years SGC Plus Gas Range Resolution Accuracy T90 Gases H2S 0 100 ppm 0 1 2 ppm 30 s CO 0 300 ppm 1 5 ppm 30 s O 0 30 vol 0 1 0 7 vol...

Page 15: ...with a damp cloth Intrinsically Safe S curit lntrins que Class I Division 1 Groups A B C and D T4 Ex ia IIC T4 Ga Class I Zone 0 AEx ia IIC T4 Ga 40 C H2S 40 C CO 35 C O Ta 50 C CAN CSA C22 2 No 60079...

Page 16: ...arranty length of coverage for products sold without an Activate Before Date shall be two 2 years beginning upon the original end user s documented date of purchase or one 1 year from the product s or...

Page 17: ...aforementioned information Please note that the defective product must be returned to GCT address which is 305 W FM 1382 Suite 540 Cedar Hill TX 75104 with provided prepaid shipping label GCT will ret...

Reviews: