SGC and SGC Plus User’s Manual

UM SGC v3.0

9 of 15

Cleaning

The detector can be cleaned with a soft damp cloth. Do not use solvents, soaps or polishes. A neutral cleaner, like Mat &

Table Top Cleaner (by ACL Staticide) may also be used.

The sensor screen may be cleaned with a soft brush under clean, warm water. Return the filter to the detector once it 

has fully dried.

Hibernation (SGC Plus Only)

When the SGC Plus is not used for extended period of time it can be hibernated to suspend the 24 month operation life 

countdown.

SGC Plus Hibernation with the GCT IR Link

1. Check that you have installed the GCT IR Link Software and the GCT IR Link USB connections are 

plugged in.

2. Click Read Device on the GCT IR Link Software.

3. Note when the detector is hibernated the event log will be cleared. It is highly recommended to save 

the event log by pressing the save event log before hibernating.

4. Click on the Hibernate Button, acknowledge the event log message.

5. Keep in front of the GCT IR Link until the “Hibernate OK” message is displayed at the bottom of the GCT 

IR Link Software.

6. Confirm the detector screen is blank.

7. If you encounter any problems please contact GCT customer support at 877-525-0808.

SGC Plus Hibernation with the Clip Dock

1. Check that the Clip Dock is turned on and USB memory is inserted.

2. The Clip Dock is capable of hibernating 4 detectors at one time: place the desired amount of detectors 

in the docking station.

3. Press and hold the bump test and calibration buttons down simultaneously for approximately 2 

seconds.

4. A successful hibernation will result in a GREEN light on for the corresponding detector number.

5. Note the event log will be automatically stored on the Clip Dock USB memory.

6. Confirm the detector screen is blank.

7. If you encounter any problems please contact GCT customer support at 877-525-0808.

Summary of Contents for Multi Gas Clip

Page 1: ...Single Gas Clip Plus Single Gas Clip User s Manual G A S C L I P T E C H C O M...

Page 2: ...n 5 Turning the Detector On 5 Turning the Detector Off 5 Performing a Self test 5 Alarms 7 Detector Maintenance 8 Bump Test Interval bUP 8 O2 Detector Calibration 8 Cleaning 9 Hibernation SGC Plus Onl...

Page 3: ...ion The CO and H2S versions of both the Single Gas Clip and SGC Plus do not require calibration for the life of the product The detector should be bump tested before each day s use with a known concen...

Page 4: ...ur l talonnage Les versions CO et H2S du Single Gas Clip et du SGC Plus ne n cessitent pas d talonnage pendant toute la dur e de vie du produit Le d tecteur doit tre test avant chaque utilisation quot...

Page 5: ...of 15 DETECTOR COMPONENTS ENTRY DESCRIPTION 1 Audible Alarm Port 2 Display 3 Alarm Bar LEDs 4 Maintenance LEDs 5 Power Button 6 Sensor 7 Infrared IR Communication Window 8 Alligator Clip with Safety...

Page 6: ...play Details The detector utilizes a special high viewing angle LCD that is designed to enhance the screen visibility In the absence of gas it displays life remaining In those cases where gas is prese...

Page 7: ...and the sensor is tested Below we outlined the most common screen configurations however if the detector has been programmed via the GCT IR Link or has been exposed to gas additional screens may appea...

Page 8: ...ctor will be reset Note while the value will be cleared on the display the value will be held in the detector s log See the event log section for further details This value can be cleared on the next...

Page 9: ...life Please discontinue use Alarm Types LOW ALARM Audible alarm One 1 slow beep every second Visual alarm One 1 slow flash every second Vibrating alarm One 1 slow vibration every second HIGH ALARM an...

Page 10: ...button press to skip the bump test If the SGC is bumped while showing GAS then it will record as a bump test in the event log instead of as an exposure If no gas is applied it will return to the norma...

Page 11: ...ve the event log by pressing the save event log before hibernating 4 Click on the Hibernate Button acknowledge the event log message 5 Keep in front of the GCT IR Link until the Hibernate OK message i...

Page 12: ...of logs you want to download a Event Logs Only b Events Partial Data Log last week of data c Events New Data Logs Only All data logs from last time you pulled the logs to current d Full Logs approxim...

Page 13: ...CALCAP Replacement calibration cap for the Single Gas Clip or the Single Gas Clip Plus Single Gas Clip Dock P N SGC DOCK portable chargeable all in one docking station in a durable portable case for a...

Page 14: ...ts Newer events replace older events Warranty Full 2 years SGC or 3 years SGC Plus Gas Range Resolution Accuracy T90 Gases H2S 0 100 ppm 0 1 2 ppm 30 s CO 0 300 ppm 1 5 ppm 30 s O 0 30 vol 0 1 0 7 vol...

Page 15: ...with a damp cloth Intrinsically Safe S curit lntrins que Class I Division 1 Groups A B C and D T4 Ex ia IIC T4 Ga Class I Zone 0 AEx ia IIC T4 Ga 40 C H2S 40 C CO 35 C O Ta 50 C CAN CSA C22 2 No 60079...

Page 16: ...arranty length of coverage for products sold without an Activate Before Date shall be two 2 years beginning upon the original end user s documented date of purchase or one 1 year from the product s or...

Page 17: ...aforementioned information Please note that the defective product must be returned to GCT address which is 305 W FM 1382 Suite 540 Cedar Hill TX 75104 with provided prepaid shipping label GCT will ret...

Reviews: