Multi Gas Clip Simple Plus User’s Manual
UM MGC-S-PLUS v1.00a
Portable gas detectors you can count on
17 of 18
Size
4.75 H x 2.75 W x 1.25 D in. (120.65 x 69.85 x 31.75 mm.)
Weight
7.675 oz. (220.2 g)
Temperature
-4˚F to +122˚F (-20˚C to +50˚C)
Humidity
5% to 95% RH (non-condensing)
Battery Life
1095 days continuous
based on an average of 1 minute/day of alarm condition
Alarms
Visual, Vibrating, Audible (minimum 95dB)
Low, High, STEL, TWA and OL (Over Limit)
LEDs
4 Red alarm bar LEDs
Yellow backlight (activated on button press)
Red backlight (activated on alarm condition)
Yellow Maintenance Notification LED
Display
Alphanumeric Liquid Crystal Display (LCD)
Logs
50 Bump Tests
50 Events
50 Calibrations
Continuous 1-second data logging (typical capacity > 2 months)
Tests
Full function self-test upon activation
Sensors, battery and circuitry tests run continuously
Ingress Protection
IP 68
Warranty
Full 3 years
Gases
Gas
Range
Resolution
Accuracy*
T
90
*
H
2
S
0 – 100 ppm
0.1
<2 ppm
<30s
CO
0 – 500 ppm
1
<5 ppm
<30s
Combustible
0 – 100% LEL
0.1
See Table 1, page 18
<30s
O
2
0 – 30% vol.
0.1
<0.7% vol.
<15s
Sensor Type
H2S, CO, O2: Single plug-in electrochemical cell
Combustible (LEL): Soldered infrared (IR)
User Options
User Message, Language, Low Alarm, High Alarm, STEL Alarm, TWA Alarm, TWA Method, TWA Interval, STEL
Interval, SAFE, Maintenance Notification, Self-test Lock, Dock Lock, Sensor Enable/Disable, Calibration
Interval, Bump Interval, Calibration Gas, %-by-volume, Latching Alarms, Life Display and Off Lock
Approvals
CSA:
Class I, Division 1, Group A, B, C, and D
Class I, Zone 0, A/Ex ia IIC T4 Ga
C22.2 No. 152-M1984
ANSI/ISA S12.13
ATEX:
SIRA 16ATEX2288
CE 0518
II 1 G Ex ia IIC T4 Ga
(-20°C ≤ Ta ≤ +50°C)
EN 60079-0:2012/A11:2013
EN 60079-11:2012
IECEx:
CSA 16.0038
Ex ia IIC T4 Ga
IEC 60079-0:2011
IEC 60079-11:2011
DETECTOR SPECIFICATIONS
*Sensor performance is dependent on many factors, including temperature, humidity, sensor age, filter cleanliness, gas
delivery, and calibration accuracy. Typical performance will be better than the given limits under most circumstances.