MGC Dock 

User’s Manual

 

V1.5 

Port able gas det ect or s y ou can count  on.  

3 of 15 

Warnings Statements/Avertisseement 

 

The Clip Dock may not support all gases. For a complete list, please contact GCT. 

 

Do not use the Clip Dock if it appears to be damaged. Inspect it before each use. 

 

Ensure the Clip Dock is used with certified calibration gas. 

 

DO NOT use expired calibration gas. Please check the expiration date located on the bottle. 

 

DO NOT operate the Clip Dock in a hazardous environment 

 

Do not expose the Clip Dock to electrical or mechanical shocks before, during, or after use. 

 

Do not allow liquids to condense and/or use high power sprays on the Clip Dock. 

 

Gas Clip Technologies recommends periodic back up of the data stored on the USB Memory. 

 

The Clip Dock contains a lithium battery that must be disposed of by a qualified recycler. Check local 

regulations for proper disposal. 

 

Warning: The battery may present a fire or chemical burn hazard if mistreated. Do not disassemble, heat 

above 100°C (212°F), or incinerate. Contact Gas Clip Technologies for replacement instructions. Use of 

another battery may present a risk of fire or explosion.  

 

DO NOT charge the instrument in temperatures above or below the specified range of 0°C to 40°C. 

 

Read the entire Clip Dock manual and follow all instructions to ensure proper use and safe installation. 

READ FIRST BEFORE OPERATION

 

Gas Clip Technologies ( GCT)   Clip Docks ar e designed t o ensur e t hat  t he port able gas 
det ect or s are able t o det ect  and alarm  t o know n t arget  gas concent rat ions. Dat a from  t hese 
t est s is st or ed t o t he onboard USB m em ory w hich can be dow nloaded for analysis at  a lat er 
t im e. I t  is y our r esponsibilit y t o pr operly use t he Clip Dock.  

 

Summary of Contents for Clip Dock

Page 1: ...User s Manual...

Page 2: ...User Operation 6 Turning the Clip Dock On and Off 6 Charging 6 Calibration Gas Installation 6 Button Usage 6 Troubleshooting Failures 7 Clip Dock Configuration 8 Clip Dock Options 8 Detector Options 8...

Page 3: ...Clip Dock contains a lithium battery that must be disposed of by a qualified recycler Check local regulations for proper disposal Warning The battery may present a fire or chemical burn hazard if mis...

Page 4: ...can be optionally housed in a rugged Pelican case or a wall mountable m etal enclosure Logs are stored on a USB flash drive which can easily be transferred to a computer for analysis Each Clip Dock co...

Page 5: ...cling charging Power LED green powered on green blinking low battery orange test in progress orange blinking No USB mem ory detected The Clip Dock will be unable to record test results if USB memory i...

Page 6: ...the bottle contains four gases so that any detector model can be tested The GCT Manager software allows you to specify the gas concentrations The default concentration is set for a mix of all gas type...

Page 7: ...eeper cavities and clear any obstructions 2 Clean the small IR com munication window located on the top of the detector 3 Verify the gas bottle is not empty 58L bottles are full at 500 PSI 116L at 100...

Page 8: ...to the PC offers faster transfers or configuration without a GCT IR Link Clip Dock Options The software program will provide a detailed description of each option if the mouse is hovered of it The Lo...

Page 9: ...fail test result Firmware and Hardware versions Location optional user string that can be configured with the GCT Manager Line 2 Bay test results for the inserted detector Bay Number 1 4 Detector Mod...

Page 10: ...on Line 2 User Message Language TWA m ethod TWA and STEL intervals SAFE mode T F Self Test Lock T F Auto Zero T F Maintenance Notification T F Dock Lock T F Latching Alarms T F The following MGC Senso...

Page 11: ...all anchors should be used For installation on brick or m etal structures the user will have to supply suitable mounting hardware A convenient wall mounting template is included to help locate the mou...

Page 12: ...ck case to slide over the screw 4 Hang the Clip Dock on the top two screws 5 Open the gas cylinder compartm ent and screw in the remaining two screws 6 To complete the installation Lead the exhaust tu...

Page 13: ...emovable USB memory Warranty Full 2 years User Options Location Gas expiration date Gas bottle lot number gas concentrations bump gas time bump sensor pass criteria calibrations bump tests hibernation...

Page 14: ...etectors you can count on 14 of 15 Contact Information Gas Clip Technologies Inc 610 Uptown Blvd Suite 4100 Cedar Hill TX 75104 Toll Free 877 525 0808 Phone 972 775 7577 Fax 972 775 2483 E mail info g...

Page 15: ...l on 1 Proper storage installation use maintenance and compliance with the user s manual instructions and any other applicable recommendations of GCT 2 The buyer promptly notifying GCT of any defect a...

Reviews: