2.2.1
POWER CORD WIRING DESCRIPTION
The PD 6500
i
(Model 11684xx) includes a standard American ground power
cord. To replace or remove plug, or to hard wire to an AC junction box, use:
Green
To Ground
Black
To line Hot
White
To line Neutral
The PD 6500
i
(Model 1168420) (INTERNATIONAL) includes a European
power cord. To replace or remove plug, or to hard wire to an AC junction
box, use:
Green/Yellow
To Ground
Brown
To line Hot
Blue
To line Neutral
2.2.2
STABILIZING
For proper stability, the PD 6500
i
should be firmly secured to the ground.
Anchor the unit to the floor, using the 1/4" mounting holes located at the base
of each panel. Position the PD 6500
i
in the desired location. Using the panels
as guides, mark and drill holes, then secure the walk-through using the
appropriate mounting devices. Tighten only to prevent rocking. DO NOT
over-tighten! As an option, stabilizer bases or adhesive floor mounting plates
are available for securing the PD 6500
i
to the floor without drilling.
2.3 PLACEMENT
PROCEDURE
1. Place assembled unit (For Assembly, See Section 2.2) in final position at
checkpoint location.
2. Connect to power source.
3. Program / Calibrate unit. (See Section 3)
4. Check for interference and troubleshoot. (See Section 5)
5. Test. (See Section 2.3.1 and 2.3.2)
CAUTION
For multiple PD 6500
i
operation (units within 25' to 100' of each other
depending on sensitivity setting), additional power requirements and
installation adjustments are necessary to avoid crosstalk (interference)
among units. (See sections 2.1.3 and 2.4 for multiple unit operation)
2.3.1
OPERATIONAL TESTING
Once the PD 6500
i
is calibrated, a SELECTED TEST OBJECT can be used
as a substitute for the forbidden object in subsequent routine verifications.
A SELECTED TEST OBJECT is an object similar in size, shape and material
composition to the smallest forbidden target and is considered an acceptable
means of verifying calibration.