Ensure the PD 6500
i
is physically stable and does not sit on top of power cord.
10.
Tighten all screws with a screwdriver.
2.2.2
POWer COrD WIrING DesCrIPTION
The PD 6500
i
(Model 11684xx) includes a standard American ground power cord. To replace
or remove plug, or to hard wire to an AC junction box, use:
Green
To Ground
Black
To line Hot
White
To line Neutral
The PD 6500
i
(Model 1168420) (INTERNATIONAL) includes a European power cord. To
replace or remove plug, or to hard wire to an AC junction box, use:
Green/Yellow
To Ground
Brown
To line Hot
Blue
To line Neutral
2.3
MULTIPLE WALK-THROUGH SITE INSTALLATION
For multiple walk-through metal detector (WTMD) operation (units within 25’ to 100’ of each
other depending on sensitivity setting), certain power requirements and installation
adjustments are necessary to avoid crosstalk (interference) among units.
There are several ways to successfully install multiple walk-throughs in close proximity with no
crosstalk. The following examples are aimed to help the installer determine an appropriate
method for eliminating crosstalk.
2.3.1
CAse 1
Description:
• All detectors are PD 6500
i
s.
(See Figure 2–7.)
• All detectors are connected to the same AC-Line Phase.
Figure 2-7
Procedure:
1.
Set all PD 6500
i
s synchronization to Sync: MASTER.
2.
Set end unit to CHANNEL 1.
3
. Set next unit to CHANNEL 2.
4.
Set next unit to CHANNEL 1.
5.
Repeat alternating pattern for channels 1 and 2 for any additional PD 6500
i
s.
Master
CH1
PD-6500
i
PD-6500
i
Master
CH2
2”
(5 cm)
min.
Master
CH1
PD-6500
i
PD-6500
i
Master
CH2
10 ft. (3m)
min.
2”
(5 cm)
min.