Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment - Guidance
Conducted
RFIEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
6 Vrms:
In ISM and amateur radio
bands between 0,15 MHz
and 80 MHz 80 % AM at
1 kHz
N/A
2
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the INDEX BPM,
including cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the trans
mitter.
Radiated RF IEC
61000-4-3
10 V/m
80 MHz - 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz - 2,7 GHz
80 % AM at 1 kHz
Recommended separation
distance:
d = 1,2
d = 1,2
80 MHz to
800 MHz
3
d = 2,3
800 MHz to
2,7 GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the trans
mitter manufacturer and
d is the recommended
separation distance in
meters (m).
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
The Index BPM is intended for use in an electromagnetic environment (for home healthcare) in which radiated
RF disturbances are controlled. The customer or the user of the Index BPM can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Index BPM as recommended in the following table, according to the maximum output
power of the communications equipment.
2
Not Applicable. The EUT is operating by battery, no supply for AC mains.
3
At 80 MHz and 800 MHz, the higher frequency range applies.
20
Device Information