Basic Safety Instructions 

2511 

Basic Safety Instructions 

2.1 

General Information 

Warning notices must be observed. Disregarding them may lead to damage to health and 
property. 
The WOB-L

®

 Piston Pumps must be operated by personnel who can detect impending dan-

gers and take action to prevent them from materialising. 
The manufacturer or authorized workshops will only service or maintain the WOB-L

®

 Piston

 

Pumps if it is accompanied by a fully completed damage report. Precise information about 
the contamination (also negative information if necessary) and thorough cleaning of the 
WOB-L

®

 Piston Pumps are legally binding parts of the contract.

 

Contaminated WOB-L

®

Piston Pumps and their individual parts must be disposed of in ac-cordance with the legal 
regulations. 
The local regulations apply in foreign countries. 

2.2 

Electricity 

The WOB-L

®

 Piston Pumps of operation mode S1 are supplied. When the location of opera-

tion mode S1 devices is changed, please note that the testing must be repeated in accor-
dance with DIN EN 0105, DIN EN 0702 and BGV A2. 
The local regulations apply in foreign countries. 
Please note the following when connecting to the electrical power supply system: 

The electrical power supply system must have a protective connector according to DIN 
VDE 0100-410 (IEC 60364-4-41). 

The protective connector must not have any breaks. 

The connecting cable must not be damaged. 

2.3 

Mechanical Systems 

Improper use can lead to injuries or material damage. Observe the following instructions: 

Only operate the pumps with hoses of the specified dimensions. 

The maximum permissible pressure of 1 bar at the suction connection must not be ex-
ceeded. 

Hazardous substances must be separated out as far as this is technically possible before 
they reach the pump. 

External mechanical stresses and vibrations must not be transmitted to the pump. Only 
use flexible laboratory hoses for connecting pumps. 

The overpressure generated at the pressure port must not exceed 3 bar. 

The pump must not be used to suck up fluids. Lay the exhaust pipe so that it slopes 
downwards, so allowing condensate to flow out of the pump. Collect the condensate and 
dispose of it in an environmentally compatible manner. 

Prevent dyes exuding. 

Maintain a space of least 20 mm between the pump and adjacent parts in order to enable 
the pump to cool. 

CAUTION ! 

Solid particles in the pumping medium impair the pumping action and can lead to 
damage. Prevent solid particles penetrating into the pump. 

Summary of Contents for Welch WOB-L 2511B-02

Page 1: ...Operation Manual EN Translation of the german original manual WOB L Piston pumps Models 2511C 02 2511B 01 2511 2017 05 16...

Page 2: ...t Prospect IL 60056 USA T 1 847 676 8800 F 1 847 677 8606 welch na gardnerdenver com www welchvacuum com We are constantly working on the further development of all our product models Reprinting or re...

Page 3: ...upply 8 3 2 Protection measure against liquids in the pump 9 3 2 1 Condensate separator Suction side 9 3 3 Areas of Application 9 3 4 Scope of Delivery 9 3 5 Accessories 9 4 Technical Data 10 5 Instal...

Page 4: ...done The user must authorize the operating personnel to do the work that has to be done The personnel must have read and understood the complete Operating Manual before using the WOB L Piston Pumps T...

Page 5: ...afety requirements Part 1 Compressors Part 2 Vacuum pumps DIN EN ISO 2151 2009 01 Acoustics Noise test code for compressors and vacuum pumps Engineering method grade 2 DIN EN 60204 12014 10 Safety of...

Page 6: ...connecting to the electrical power supply system The electrical power supply system must have a protective connector according to DIN VDE 0100 410 IEC 60364 4 41 The protective connector must not have...

Page 7: ...recommended for pumping acidic basic or organic va pors Poisonous Gases Use a condensate separator eg Woulff bottle when pumping poisonous or harmful gases Prevent such substances from leaking out of...

Page 8: ...e disconnection devices current output Motor is provided with a thermal overload protection ex works protecting the vacuum pump from damage or destruction respectively WARNING Should the user change t...

Page 9: ...hemical laboratories in trade and industry Use for vacuum filtration vacuum drying and other vacuum technology applications 3 4 Scope of Delivery The scope of delivery is specified in the supply contr...

Page 10: ...er No for WOB L Piston pump 230V 50Hz inclusive mains connection cable with plug CEE and UK 2511C 02 WOB L Piston pump 115V 60Hz inclusive mains connection cable with plug US 2511B 01 The information...

Page 11: ...the cooling of the pump is guaranteed see chapter 2 3 Note that on the installation location no moisture acts on the pump 5 2 2 Connecting 5 2 2 1 Electrical Connection Before the electrical connectio...

Page 12: ...he pump 5 3 2 Decommissioning If vacuum applications with special steam load the pump needs to run after the process with an open vacuum port about 2 minutes and rinsed with atmospheric air This measu...

Page 13: ...ilding up on the surface of the pump Check the electrical and vacuum connections daily 6 2 Maintenance by the Manufacturer Repairs and maintenance going beyond the extent of the work described in chap...

Page 14: ...rvice workshop Exchange Pump does not generate a vacuum or overpressure Connected apparatus and or connecting elements leaking User or Service workshop Identify and seal the leak replace the seals and...

Page 15: ...ssessment and risk reduction S curit des machines Principes g n raux pour l valuation des risques et la r duction des risques Seguridad de las m quinas Principios generales para el dise o Evaluaci n d...

Reviews: