Operation
36
115051
In the PSI pressure range, the pressure values for the curve are multiplied by 100 to remain
with integers.
The curve can be printed out by the system printer via the "File / Print curve" menu.
The buttons largely correspond to the buttons on the controller. However, the buttons cannot
be held down.
In active mode, the
AIR
button can be used with the mouse like on the controller.
Pressing the mouse button opens the valve, releasing it closes the valve again.
However, this only works if the mouse pointer is still on the
AIR
button.
The
Up
and
Down
arrow buttons change the temporary setpoint in active mode.
The
OK
button only works in active mode, and sets the current pressure as the new set
value.
The
Start
button starts or ends in the current operating mode.
The
Pmin
button functions the same as without a controller.
The following menu items can be selected from the menu bar:
File
Measure!
Settings!
Calibrate!
Factory settings!
6.
8
.1
Menu item: „File“
File
Save settings
Saving the Settings dialog settings in a file
Load settings
Loading the Settings dialog settings from a file
Print settings
Printing out an image of the Settings dialog on the standard printer
Save memory
Saving the pressure data in a file
Load memory
Load the print data from a file
Print curve
Printing out the current pressure curve on the standard printer
Password
Changing the basic setting of the controller is password-protected.
Release is given by inputting the date as a hex number.
Example :
15.06.2018 = 150618 = 24C5A
(Required for activating the "Calibrate" and "Factory settings" menus and the direct input
of pressure offset und pressure factor in the „Calibrate“ Dialog)
Info
Information about the program version
Exit <ESC>
Exit program