background image

EG - Konformitätserklärung 

EC Declaration of Conformity / CE Déclaration de Conformité   

DIN EN ISO / IEC 17050 

 

 

(de) 
 
Hiermit erklären wir 

 

 

Gardner Denver Thomas GmbH 
Am Vogelherd 20 
98693 Ilmenau 
Germany 

 

 

T +49 3677 604 0 
F +49 3677 604 131 

[email protected] 
www.welchvacuum.com 

 

unter eigener Verantwortung, dass nachstehendes Produkt aufgrund seiner Konzipierung und Bauart sowie in den von uns in Verkehr gebrachten Unter-
lagen den nachfolgend aufgeführten EG-Richtlinien und Normen entspricht.  
Bei einer nicht mit uns abgestimmten Änderung des Produkts verliert diese Erklärung ihre Gültigkeit.
 

(en)  We (Gardner Denver Thomas GmbH) herewith declare under our sole responsibility that the product described below is in accordance with the following 
         Directives standards and other technical specifications regarding design and version when delivered from our factory. 
         This declaration becomes invalid whenever the product has been modified without our consent. 

(fr)   Nous (Gardner Denver Thomas GmbH) certifions par la présente, que le produit décrit ci-après est conforme, tant dans sa conception que dans sa réalisation, 
        aux normes de sécurité et d'hygiène exigées par les standards de la CE. 
        En cas de modification du produit sans notre accord, cette déclaration devient caduque.
 

 

Bezeichnung des Produkts (Pumpen / Pumpstände) 
Description of product (pumps / pump systems) 
Description du produit (pompes / pompe systèmes) 

chemvac Kombipumpstände /  

chemvac Combination Pump Systems / Pompes combinée chemvac 

P 6 Z - 101,   P 12 Z - 301,   P 23 Z - 301 

Artikel-Nr. / Fabrication No. / No. de fabrication 

109030, 109030-01, 109030-03, 

109031, 109031-03, 109032, 109032-03 

Baujahr / Year of manufacture / Annee de fabrication 

2017 

 

 

Das Produkt entspricht folgenden Richtlinien und Normen: / The product is in conformity with the following Directives and stan-
dards: / Le produit est conforme aux directives et standards suivants: 

2006/42/EG   

Maschinenrichtlinie / EC machinery directive / directive CE sur les machines (17.05.2006) 

 

2014/34/EU   

ATEX-Richtlinie für Verwendungen in explosionsgefährdeten Bereichen, Anhang III / ATEX Guideline for use in potentially 
explosive atmospheres, Appendix III / ATEX Directive for applications in hazardous areas, Annex III 

2014/30/EU   

Elektromagnetische Verträglichkeit / EC Electromagnetic Compatibility Directive / Directive CE relative à la compatibilité électro-
magnétique
 

2011/65/EU   

Gefährliche Stoffe in Elektro- und Elektronikgeräten (RoHS II) / Dangerous materials in electrical and electronics devices 
(RoHS II) / Substances dangereuses dans les appareils électriques et électroniques (RoHS II) 

2012/19/EU   

Elektro- und Elektronik - Altgeräte (WEEE) / Electrical and electronics - old devices (WEEE) / Électro et électronique - appareils de 
contralto (WEEE)
 

China – RoHS II 

Umweltschutzgesetz – China 2016-01 / Environment protection law / Loi sur la protection de environnement  

 

Angewandte harmonisierte Normen: / Applied harmonized standards: / Standards appliques et harmonises: 

 

DIN EN 1127-1: 
2011-10
 

Explosionsfähige Atmosphären – Explosionsschutz - Teil 1: Grundlagen und Methodik /  
Explosive atmospheres - Explosion prevention and protection - part 1: Basic concepts and methodology /  
Atmosphères explosives - Protection contre les explosions - partie 1 : prescriptions et méthodologie 

 

DIN EN 13463-1: 
2009-07
 

Nicht-elektrische Geräte für den Einsatz in explosionsgefährdeten Bereichen - Teil 1: Grundlagen und Anforderungen / Non-
electrical equipment for use in potentially explosive atmospheres - part 1: Basic method and requirements / Appareils non électriques 
destinés à être utilisés en atmosphères explosibles - partie 1 : prescriptions et méthodologie
 

 

DIN EN 13463-5: 
2011-10
 

Nicht-elektrische Geräte für den Einsatz in explosionsgefährdeten Bereichen -  Teil 5: Schutz durch konstruktive Sicherheit 
‚c’ / 
Non-electrical equipment for use in potentially explosive atmospheres - part 5: Protection by constructional safety 'c' / Appareils 
non électriques destinés à être utilisés en atmosphères explosibles - partie 5 : protection par sécurité de construction « c »
 

DIN EN ISO 12100: 
2011-03  
 

Sicherheit von Maschinen - Allgemeine Gestaltungsleitsätze Risikobeurteilung und Risikominderung /  
Safety of machinery - General principles for design - Risk assessment and risk reduction /

  

Sécurité des machines - / Principes généraux pour l'évaluation des risques et la réduction des risques 

DIN EN ISO 13857: 
2008-06  
 

Sicherheit von Maschinen - Sicherheitsabstände gegen das Erreichen von Gefährdungsbereichen mit den oberen und 
unteren Gliedmaßen
 / Safety of machinery - Safety distances to prevent hazard zones being reached by upper and lower limbs / 
Sécurité des machines - Distances de sécurité empêchant les membres supérieurs et inférieurs d'atteindre les zones dangereuses 

DIN EN 1012-2: 
2011-12  
 

Kompressoren und Vakuumpumpen - Sicherheitsanforderungen - Teil 2: Vakuumpumpen /  
Compressors and vacuum pumps - Safety requirements - part 2: Vacuum pumps /  
Compresseurs et pompes à vide - Exigences de sécurité - partie 2: pompes à vide 

DIN EN ISO 2151: 
2009-01
 

Akustik - Geräuschmessnorm für Kompressoren und Vakuumpumpen - Verfahren der Genauigkeitsklasse 2 /  
Acoustics - Noise test code for compressors and vacuum pumps – Engineering method (grade 2) / Acoustique - norme de mesure des 
émissions pour les compresseurs et les pompes à vide - Procédé de classe de précision 2
 

DIN EN 60204-1: 
2014-10
 

Sicherheit von Maschinen - Elektrische Ausrüstung von Maschinen - Teil 1: Allgemeine Anforderungen /  
Safety of machinery - Electrical equipment of machines - part 1: General requirements /  
Sécurité des machines - Equipement électrique des machines - partie 1: Prescriptions générales 

EN 61000-6-2: 
2011-06
 

Elektromagnetische Verträglichkeit (EMV) - Teil 6-2: Fachgrundnormen - Störfestigkeit für Industriebereiche /  
Electromagnetic compatibility (EMC) - part 6-2: Generic standards - Immunity for industrial environments / Compatibilité électromag-
nétique (EMV) - partie 6-2: Normes génériques - Immunité pour les environnements industriels
 

EN 61000-6-4: 
2011-09
 

Elektromagnetische Verträglichkeit (EMV) - Teil 6-4: Fachgrundnormen - Störaussendung für Industriebereiche / Electromag-
netic compatibility (EMC) - part 6-4: Generic standards - Emission standard for industrial environments environments / Compatibilité 
électromagnétique - partie 6-4: Normes génériques - Emissions de parasites pour les activités industrielles
 

DIN EN 50110-1: 
2014-02
 

Betrieb von elektrischen Anlagen / 

Operation of electrical installations / 

Fonctionnement des installations électriques 

DIN EN 61010-
1/A1:2015-04 

Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 1: Allgemeine Anforderungen / 

Safety 

requirements for electrical equipment for measurement, control and laboratory use - part 1: General requirements / 

Consignes de 

sécurité pour les appareils électriques de mesure, de commande, de régulation ou de laboratoire - partie 1: Prescriptions générales 

 

Datum / Data 

2017-02-16 

Qualitätsbeauftragter / Quality representative / Délégué de qualité 

Name / Name / Nom 

Gerd Reinhardt 

 

Produktmanager / Product manager / Directeur de produit 

Name / Name / Nom 

Oliver Fickert 

 

 

Summary of Contents for WELCH chemvac P 6 Z

Page 1: ...Operation Manual EN Translation of the german original manual chemvac Combination Pump Systems Models P 6 Z chemvac P 12 Z chemvac P 23 Z chemvac 109030 2017 03 17...

Page 2: ...ritten permission of Co Gardner Denver Thomas GmbH All rights under the copyright laws are expressly reserved by Co Gardner Denver Thomas GmbH We reserve the right to make changes and amendments Gardn...

Page 3: ...8 6 2 Maintenance Performed by the User 18 6 2 1 Maintenance of the rotary vane pump 19 6 2 1 1 Oil check 19 6 2 1 2 Oil change 19 6 2 1 3 Changing the LABOVAC oil type 21 6 2 2 Maintenance of the dia...

Page 4: ...ersonnel for combination pump systems The personnel operating and maintaining the rotary vane pumps must have the technical competence required to perform the work that has to be done The user must au...

Page 5: ...O 2151 2009 01 Acoustics Noise test code for compressors and vacuum pumps Engineering method grade 2 DIN EN 60204 1 2014 10 Safety of machinery Electrical equipment of machines Part 1 General requirem...

Page 6: ...sed of in accordance with the legal regulations The local regulations apply in foreign countries 2 2 Electricity The chemvac Combination Pump Systems are supplied for operating mode S1 Please note tha...

Page 7: ...the pumps and adjacent parts in order to enable the pumps to cool WARNING Do not block the exhaust port Remove any exhaust caps or plugs Otherwise pres sure will build up with potential of bursting ha...

Page 8: ...is not permitted to pump gases that are combustible or prone to explosion Explosive gases The chemvac Combination Pump Systems is not certified according to ATEX guidelines 2014 34 EU Aggressive gase...

Page 9: ...ke pressures and the pressure back up which this causes between the pumps is compensated for by the pressure control valve 4 integrated into the valve block 3 The separator AKD 3 catches condensates f...

Page 10: ...either do not or only slightly condense because of the pump operating temperature and the under pressure created in the oil casing via the diaphragm pump The under pressure is indicated by the manomet...

Page 11: ...pumping out vapours including those soluble in oil ceramic sintering plants with aggressive by products and plasma etching plants freeze drying Special designs Special chemvac combination pump systems...

Page 12: ...pes Order no B T H dimensions in mm P 6 Z 101 109030 109030 01 109030 03 500 330 340 P 12 Z 301 109031 109031 03 590 345 420 P 23 Z 301 109032 109031 03 590 345 420 Fig 3 Dimensions 4 2 Intake Pressur...

Page 13: ...2151 dB A 48 50 50 Water vapour tolerance mbar 70 80 45 Oil filling Type LABOVAC 14 enclosed 1 litre bottle ml 550 1000 820 Voltage V 100 120 200 240 switchable Frequency Hz 50 60 Power complete W 400...

Page 14: ...tion pressure port small flange DN 16 KF DN 25 KF DN 25 KF Ambient temperature C 20 to 40 Max operating gas temperature 40 Noise level DIN EN ISO 2151 dB A 48 50 50 Water vapour tolerance mbar 60 35 3...

Page 15: ...perating mode S 1 Type of protection DIN EN 60529 IP 54 Motor Class of insulation DIN EN 600034 1 BLDC F 160 C Dimensions W D H mm 195 235 145 230 265 169 Weight kg 6 5 11 2 Order numbers for Diaphrag...

Page 16: ...e chemvac combination pump system without any oil filling Therefore we enclose a separate bottle with one litre of our pumping oil for rotary vane pumps type LABOVAC 14 which you need to fill into the...

Page 17: ...t formation of foam CAUTION For optimum operation of the pump stand the user may not change the factory Bubbler setting see also sign at the Bubbler That is only the manufacturer and or one of it auth...

Page 18: ...the disposal of haz ardous substances Check that the filter cartridge is in good working order The filter car tridge must be exchanged if the in built valve is actuated during operation Check that th...

Page 19: ...4 lower marking MIN 5 upper marking MAX Fig 5 Oil level control 6 2 1 1 Oil check CAUTION The condition and quality of the pump oil have a substantial effect upon the per formance and operational rea...

Page 20: ...with the valid environmental protection regulations Filling up with oil Remove the oil filler plug Pour in the oil until it reaches the upper mark Screw in the oil filling plug together with the seal...

Page 21: ...thetic oils like LABOVAC 11 and LABOVAC 14 with mineral oils like LABOVAC 10 or LABOVAC 12S will be executed in the same way The synthetic oil LABOVAC 13 is an exception which CANNOT be mixed or excha...

Page 22: ...ts if necessary Wear protective gloves Parts must be renewed at the intervals stated in this Operating Manual or as speci fied by the user internally Do not clean with compressed air 6 2 2 2 Assembly...

Page 23: ...ssembly MPC 301 Z 1 Place the pump so that the diaphragm 7 is lying in a horizontal position 2 Use the size 3 pin type face wrench to tighten the pressure disc 9 the diaphragm 7 and the strain washer...

Page 24: ...pump 6 3 1 Servicing WARNING During repair or maintenance work which could endanger people because of moving or electrically live components the vacuum pump must be made safe by removing the mains sup...

Page 25: ...mp defective Repair and or exchange Diaphragm pump defective Oil is heavily contami nated by condensate Change oil Pressure in the oil casing of the rotary vane pump is too high outside the green rang...

Page 26: ...r no 302081 for P 12 Z P 23 Z Order no 302082 consisting of Piece Order no Piece Order no Radial shaft seal FKM 18 x 24 x 3 1 829417 Shaft seal FKM 18 x 28 x 7 BAUMX7 1 829413 Shaft seal FKM 18 x 30 x...

Page 27: ...ug 1 824104 1 824104 1 824104 O Ring FKM 48 x 2 1 829265 1 829265 1 829265 Rotary vane oil pump 2 300149 01 2 300072 2 300072 Rotary vane high stage 2 300893 2 300940 03 2 300916 12 Rotary vane pre st...

Page 28: ...Overview of spare parts 28 109030 8 4 Complete view chemvac combination pump system Fig 8 Complete view chemvac combination pump system P 6 Z 101...

Page 29: ...03 109502 03 109502 03 Threaded elbow joint PVDF 8 1 8 1 829936 1 829936 1 829936 1 Straight threaded joint PVDF 8 1 8 1 829919 829919 829919 O ring FKM 30 x 2 1 829291 829291 829291 O ring FKM 18 x...

Page 30: ...Overview of spare parts 30 109030 8 5 Spare part view P 6 Z Fig 9 Exploded view P 6 Z...

Page 31: ...1 300931 04 19 Pump casing pre stage 1 300931 05 20 Rotor pre stage 1 300931 10 21 Rotary vane pre stage 2 300876 22 Rotary vane oil pump 2 300149 01 23 Oil pump 1 109509 02 24 Control piston 1 300368...

Page 32: ...Overview of spare parts 32 109030 8 6 Spare part view P 12 Z P 23 Z Fig 10 Exploded view P 12 Z P 23 Z...

Page 33: ...19 Pump casing pre stage 1 300916 03 20 Rotor pre stage 1 300916 09 21 Rotary vane pre stage 2 300916 11 22 Rotary vane oil pump 2 300072 23 Oil pump 1 109510 02 24 Cover 1 300074 25 Compressing piec...

Page 34: ...4 19 Pump casing pre stage 1 300916 03 20 Rotor pre stage 1 300916 09 21 Rotary vane pre stage 2 300916 11 22 Rotary vane oil pump 2 300072 23 Oil pump 1 109510 02 24 Cover 1 300074 25 Compressing pie...

Page 35: ...motor 115 V 1 410310 01 Drive complete consisting of position 3 5 1 400763 01 400763 01 3 Eccentric 1 400770 400770 4 Piston rod with ball bearing 2 400771 01 400771 01 5 Close tolerance spacer 20 x 2...

Page 36: ...acer 25 x 35 x 1 4 824957 1 6 Piston rod with ball bearing 2 400647 01 7 Mass balance 1 400678 8 Alternating current motor 1 826420 01 9 Rubber metal pad 4 829139 10 Rubber pad 1 400785 01 11 Pressure...

Page 37: ...Product Code 201560250510 601609 60 1 2 RELEVANT IDENTIFIED USES OF THE SUBSTANCE OR MIXTURE AND USES ADVISED AGAINST Intended Use Compressor oil Uses advised against None unless specified elsewhere...

Page 38: ...rpose than the intended use in Section 1 without expert advice Health studies have shown that chemical exposure may cause potential human health risks which may vary from person to person SECTION 3 CO...

Page 39: ...ing specific and immediate medical treatment available in the workplace is not expected SECTION 5 FIRE FIGHTING MEASURES 5 1 EXTINGUISHING MEDIA Suitable Extinguishing Media Use water fog foam dry che...

Page 40: ...pill wave and current direction and speed may greatly influence the appropriate action to be taken For this reason local experts should be consulted Note Local regulations may prescribe or limit actio...

Page 41: ...tilation Information on the selection of protective equipment for use with this material as provided below is based upon intended normal usage Respiratory Protection If engineering controls do not mai...

Page 42: ...annot be cleaned Practice good housekeeping ENVIRONMENTAL CONTROLS Comply with applicable environmental regulations limiting discharge to air water and soil Protect the environment by applying appropr...

Page 43: ...temperatures SECTION 11 TOXICOLOGICAL INFORMATION 11 1 INFORMATION ON TOXICOLOGICAL EFFECTS Hazard Class Conclusion Remarks Inhalation Acute Toxicity No end point data for material Minimally Toxic Ba...

Page 44: ...Repeated Exposure No end point data for material Not expected to cause organ damage from prolonged or repeated exposure Based on assessment of the components OTHER INFORMATION Contains Synthetic base...

Page 45: ...oper waste disposal code s This material is considered as hazardous waste pursuant to Directive 91 689 EEC on hazardous waste and subject to the provisions of that Directive unless Article 1 5 of that...

Page 46: ...1272 2008 on classification labelling and packaging of substances and mixtures and amendments thereto Refer to the relevant EU national regulation for details of any actions or restrictions required b...

Page 47: ...thal Concentration LD Lethal Dose LL Lethal Loading EC Effective Concentration EL Effective Loading NOEC No Observable Effect Concentration NOELR No Observable Effect Loading Rate THIS SAFETY DATA SHE...

Page 48: ...Revision Date 2014 06 05 Page 12 of 12 _____________________________________________________________________________________________________________________ DGN 2007484XGB 547849 ANNEX Annex not requ...

Page 49: ...equipment for use in potentially explosive atmospheres part 1 Basic method and requirements Appareils non lectriques destin s tre utilis s en atmosph res explosibles partie 1 prescriptions et m thodo...

Reviews: