background image

Basic Safety Instructions 

109030 

Basic Safety Instructions 

2.1 

General Information 

WARNING ! 

Warning notices must be observed. Disregarding them may lead to damage to health 
and property. 

The chemvac Combination Pump Systems must be operated by personnel who can detect 
impending dangers and take action to prevent them from materialising. 
The user/operator is responsible for correct installation and safe operation. 
The manufacturer or authorized authorised workshops will only service or maintain the 
chemvac combination pump systems if it is accompanied by a fully completed damage re-
port. Precise information about the contamination (also negative information if necessary) 
and thorough cleaning of the laboratory vacuum system are legally binding parts of the con-
tract. 
Contaminated chemvac Combination Pump Systems and their individual parts must be dis-
posed of in accordance with the legal regulations. 
The local regulations apply in foreign countries. 

2.2 

Electricity 

The chemvac Combination Pump Systems are supplied for operating mode S1. Please note 
that the testing must be repeated in accordance with DIN EN 0105, DIN EN 0702 and BGV 
A2 in case of portable devices. 
The local regulations apply in foreign countries. 

Please note the following when connecting to the electrical power supply system: 

The electrical power supply system must have a protective connector according to DIN 
VDE 0100-410 (IEC 60364-4-41). 

The protective connector must not have any breaks. 

The connecting cable must not be damaged. 

2.3 

High Temperatures 

The pumps of the chemvac combination pump system can become heated by the tempera-
ture of the gas being pumped and the heat of compression.  
Prevent the following maximum permissible temperatures from being exceeded. 

+ 40 °C for the environment, and  

+ 20 °C for the gas to be pumped. 

The motor for single phase alternating current is protected against overload by an integrated 
motor protection switch. 

Summary of Contents for WELCH chemvac P 6 Z

Page 1: ...Operation Manual EN Translation of the german original manual chemvac Combination Pump Systems Models P 6 Z chemvac P 12 Z chemvac P 23 Z chemvac 109030 2017 03 17...

Page 2: ...ritten permission of Co Gardner Denver Thomas GmbH All rights under the copyright laws are expressly reserved by Co Gardner Denver Thomas GmbH We reserve the right to make changes and amendments Gardn...

Page 3: ...8 6 2 Maintenance Performed by the User 18 6 2 1 Maintenance of the rotary vane pump 19 6 2 1 1 Oil check 19 6 2 1 2 Oil change 19 6 2 1 3 Changing the LABOVAC oil type 21 6 2 2 Maintenance of the dia...

Page 4: ...ersonnel for combination pump systems The personnel operating and maintaining the rotary vane pumps must have the technical competence required to perform the work that has to be done The user must au...

Page 5: ...O 2151 2009 01 Acoustics Noise test code for compressors and vacuum pumps Engineering method grade 2 DIN EN 60204 1 2014 10 Safety of machinery Electrical equipment of machines Part 1 General requirem...

Page 6: ...sed of in accordance with the legal regulations The local regulations apply in foreign countries 2 2 Electricity The chemvac Combination Pump Systems are supplied for operating mode S1 Please note tha...

Page 7: ...the pumps and adjacent parts in order to enable the pumps to cool WARNING Do not block the exhaust port Remove any exhaust caps or plugs Otherwise pres sure will build up with potential of bursting ha...

Page 8: ...is not permitted to pump gases that are combustible or prone to explosion Explosive gases The chemvac Combination Pump Systems is not certified according to ATEX guidelines 2014 34 EU Aggressive gase...

Page 9: ...ke pressures and the pressure back up which this causes between the pumps is compensated for by the pressure control valve 4 integrated into the valve block 3 The separator AKD 3 catches condensates f...

Page 10: ...either do not or only slightly condense because of the pump operating temperature and the under pressure created in the oil casing via the diaphragm pump The under pressure is indicated by the manomet...

Page 11: ...pumping out vapours including those soluble in oil ceramic sintering plants with aggressive by products and plasma etching plants freeze drying Special designs Special chemvac combination pump systems...

Page 12: ...pes Order no B T H dimensions in mm P 6 Z 101 109030 109030 01 109030 03 500 330 340 P 12 Z 301 109031 109031 03 590 345 420 P 23 Z 301 109032 109031 03 590 345 420 Fig 3 Dimensions 4 2 Intake Pressur...

Page 13: ...2151 dB A 48 50 50 Water vapour tolerance mbar 70 80 45 Oil filling Type LABOVAC 14 enclosed 1 litre bottle ml 550 1000 820 Voltage V 100 120 200 240 switchable Frequency Hz 50 60 Power complete W 400...

Page 14: ...tion pressure port small flange DN 16 KF DN 25 KF DN 25 KF Ambient temperature C 20 to 40 Max operating gas temperature 40 Noise level DIN EN ISO 2151 dB A 48 50 50 Water vapour tolerance mbar 60 35 3...

Page 15: ...perating mode S 1 Type of protection DIN EN 60529 IP 54 Motor Class of insulation DIN EN 600034 1 BLDC F 160 C Dimensions W D H mm 195 235 145 230 265 169 Weight kg 6 5 11 2 Order numbers for Diaphrag...

Page 16: ...e chemvac combination pump system without any oil filling Therefore we enclose a separate bottle with one litre of our pumping oil for rotary vane pumps type LABOVAC 14 which you need to fill into the...

Page 17: ...t formation of foam CAUTION For optimum operation of the pump stand the user may not change the factory Bubbler setting see also sign at the Bubbler That is only the manufacturer and or one of it auth...

Page 18: ...the disposal of haz ardous substances Check that the filter cartridge is in good working order The filter car tridge must be exchanged if the in built valve is actuated during operation Check that th...

Page 19: ...4 lower marking MIN 5 upper marking MAX Fig 5 Oil level control 6 2 1 1 Oil check CAUTION The condition and quality of the pump oil have a substantial effect upon the per formance and operational rea...

Page 20: ...with the valid environmental protection regulations Filling up with oil Remove the oil filler plug Pour in the oil until it reaches the upper mark Screw in the oil filling plug together with the seal...

Page 21: ...thetic oils like LABOVAC 11 and LABOVAC 14 with mineral oils like LABOVAC 10 or LABOVAC 12S will be executed in the same way The synthetic oil LABOVAC 13 is an exception which CANNOT be mixed or excha...

Page 22: ...ts if necessary Wear protective gloves Parts must be renewed at the intervals stated in this Operating Manual or as speci fied by the user internally Do not clean with compressed air 6 2 2 2 Assembly...

Page 23: ...ssembly MPC 301 Z 1 Place the pump so that the diaphragm 7 is lying in a horizontal position 2 Use the size 3 pin type face wrench to tighten the pressure disc 9 the diaphragm 7 and the strain washer...

Page 24: ...pump 6 3 1 Servicing WARNING During repair or maintenance work which could endanger people because of moving or electrically live components the vacuum pump must be made safe by removing the mains sup...

Page 25: ...mp defective Repair and or exchange Diaphragm pump defective Oil is heavily contami nated by condensate Change oil Pressure in the oil casing of the rotary vane pump is too high outside the green rang...

Page 26: ...r no 302081 for P 12 Z P 23 Z Order no 302082 consisting of Piece Order no Piece Order no Radial shaft seal FKM 18 x 24 x 3 1 829417 Shaft seal FKM 18 x 28 x 7 BAUMX7 1 829413 Shaft seal FKM 18 x 30 x...

Page 27: ...ug 1 824104 1 824104 1 824104 O Ring FKM 48 x 2 1 829265 1 829265 1 829265 Rotary vane oil pump 2 300149 01 2 300072 2 300072 Rotary vane high stage 2 300893 2 300940 03 2 300916 12 Rotary vane pre st...

Page 28: ...Overview of spare parts 28 109030 8 4 Complete view chemvac combination pump system Fig 8 Complete view chemvac combination pump system P 6 Z 101...

Page 29: ...03 109502 03 109502 03 Threaded elbow joint PVDF 8 1 8 1 829936 1 829936 1 829936 1 Straight threaded joint PVDF 8 1 8 1 829919 829919 829919 O ring FKM 30 x 2 1 829291 829291 829291 O ring FKM 18 x...

Page 30: ...Overview of spare parts 30 109030 8 5 Spare part view P 6 Z Fig 9 Exploded view P 6 Z...

Page 31: ...1 300931 04 19 Pump casing pre stage 1 300931 05 20 Rotor pre stage 1 300931 10 21 Rotary vane pre stage 2 300876 22 Rotary vane oil pump 2 300149 01 23 Oil pump 1 109509 02 24 Control piston 1 300368...

Page 32: ...Overview of spare parts 32 109030 8 6 Spare part view P 12 Z P 23 Z Fig 10 Exploded view P 12 Z P 23 Z...

Page 33: ...19 Pump casing pre stage 1 300916 03 20 Rotor pre stage 1 300916 09 21 Rotary vane pre stage 2 300916 11 22 Rotary vane oil pump 2 300072 23 Oil pump 1 109510 02 24 Cover 1 300074 25 Compressing piec...

Page 34: ...4 19 Pump casing pre stage 1 300916 03 20 Rotor pre stage 1 300916 09 21 Rotary vane pre stage 2 300916 11 22 Rotary vane oil pump 2 300072 23 Oil pump 1 109510 02 24 Cover 1 300074 25 Compressing pie...

Page 35: ...motor 115 V 1 410310 01 Drive complete consisting of position 3 5 1 400763 01 400763 01 3 Eccentric 1 400770 400770 4 Piston rod with ball bearing 2 400771 01 400771 01 5 Close tolerance spacer 20 x 2...

Page 36: ...acer 25 x 35 x 1 4 824957 1 6 Piston rod with ball bearing 2 400647 01 7 Mass balance 1 400678 8 Alternating current motor 1 826420 01 9 Rubber metal pad 4 829139 10 Rubber pad 1 400785 01 11 Pressure...

Page 37: ...Product Code 201560250510 601609 60 1 2 RELEVANT IDENTIFIED USES OF THE SUBSTANCE OR MIXTURE AND USES ADVISED AGAINST Intended Use Compressor oil Uses advised against None unless specified elsewhere...

Page 38: ...rpose than the intended use in Section 1 without expert advice Health studies have shown that chemical exposure may cause potential human health risks which may vary from person to person SECTION 3 CO...

Page 39: ...ing specific and immediate medical treatment available in the workplace is not expected SECTION 5 FIRE FIGHTING MEASURES 5 1 EXTINGUISHING MEDIA Suitable Extinguishing Media Use water fog foam dry che...

Page 40: ...pill wave and current direction and speed may greatly influence the appropriate action to be taken For this reason local experts should be consulted Note Local regulations may prescribe or limit actio...

Page 41: ...tilation Information on the selection of protective equipment for use with this material as provided below is based upon intended normal usage Respiratory Protection If engineering controls do not mai...

Page 42: ...annot be cleaned Practice good housekeeping ENVIRONMENTAL CONTROLS Comply with applicable environmental regulations limiting discharge to air water and soil Protect the environment by applying appropr...

Page 43: ...temperatures SECTION 11 TOXICOLOGICAL INFORMATION 11 1 INFORMATION ON TOXICOLOGICAL EFFECTS Hazard Class Conclusion Remarks Inhalation Acute Toxicity No end point data for material Minimally Toxic Ba...

Page 44: ...Repeated Exposure No end point data for material Not expected to cause organ damage from prolonged or repeated exposure Based on assessment of the components OTHER INFORMATION Contains Synthetic base...

Page 45: ...oper waste disposal code s This material is considered as hazardous waste pursuant to Directive 91 689 EEC on hazardous waste and subject to the provisions of that Directive unless Article 1 5 of that...

Page 46: ...1272 2008 on classification labelling and packaging of substances and mixtures and amendments thereto Refer to the relevant EU national regulation for details of any actions or restrictions required b...

Page 47: ...thal Concentration LD Lethal Dose LL Lethal Loading EC Effective Concentration EL Effective Loading NOEC No Observable Effect Concentration NOELR No Observable Effect Loading Rate THIS SAFETY DATA SHE...

Page 48: ...Revision Date 2014 06 05 Page 12 of 12 _____________________________________________________________________________________________________________________ DGN 2007484XGB 547849 ANNEX Annex not requ...

Page 49: ...equipment for use in potentially explosive atmospheres part 1 Basic method and requirements Appareils non lectriques destin s tre utilis s en atmosph res explosibles partie 1 prescriptions et m thodo...

Reviews: