23
If a command is invalid, one of the following error codes is echoed as well as
being displayed on the LEDs.
E33 Unrecognised command or parameters.
E44 Illegal mode or channel number in command.
Commands take effect immediately they are issued. Because of this, if the
settings are being changed from high current Pulsed Output mode to a
Continuous Output mode, it is advisable to set the current values to 0 using the
RC command before changing mode. This prevents overload conditions
occurring.
Any changes made using RS232 commands are not saved permanently
until the AW command has been issued.
10.1.1
General Commands
AW
Save the settings to memory.
Syntax:
AW
Once the settings are saved to memory they are then retained when the
unit is switched off. If this is not done, changes to the settings are volatile,
and if the unit is switched off they revert to those in force when the last
AW command was issued.
ST
Report the current settings for the outputs
Syntax:
ST
The 1
st
and 2
nd
lines of the response report the output settings. The format
is either:
f
M
m
E
e
V
a,b,c,d
For Constant or Selected Mode
Or:
f
M
m
E
e
C
a
D
del
P
p
For Pulsed Output Mode
Where:
f
output channel (1 or 2)
m is one of:
0
Continuous Output
1
Trigger on Input 1
2
Trigger on Input 2
3
Trigger on both inputs
4
Select on Input 1
5
Select on Input 2
6
Select using both inputs
a
Output current 1