GARANT Digital dial indicator HCT
13. Technical data
13.1.
DIAL INDICATOR
Data
12.5 mm
25 mm
Measuring range
12.5 mm
25 mm
Error limit
(10 μm resolution)
10 μm (± 1 digit)
10 μm (± 1 digit)
Error limit
(1 μm resolution)
3 μm
4 μm
Error limit
(0.1 μm resolution)
1.8 μm
2.2 μm
Repeatability
2 μm
2 μm
Weight
119 g
123 g
Gauging force (standard)
0.8-1.4 N
0.8-1.7 N
Maximum
adjustment speed
1.7 m/s
Number of measurements per second
Measurement: 10 /s
MIN/MAX mode: 20 /s
Unit of measurement
mm / inch
Maximum for preset
(10 μm resolution)
±9999.99 mm / ±399.9995 IN
Maximum for preset (1 μm resolution)
±999.999 mm / ±39.99995 IN
Maximum for preset (0.1 μm resolution)
±99.9999 mm / ±3.999995 IN
Graduation for scale
Minimum 0.2 μm – maximum 1 mm / min. 0.01 thou - 0.05 IN
Measuring range for scale
Minimum ±5 μm – maximum ±25 mm / min. ± 0.25 thou – max. ± 1.25 IN
Measuring system
Sylvac inductive system (patented)
Supply
1 x 3 V lithium battery, type CR2032, capacity 220 mAh
Data output
RS232/USB compatible
Operating temperature
+5 °C to +40 °C
Storage temperature
-10 °C to +60 °C
Electromagnetic
compatibility
to EN 61326-1
IP specification
to IEC60529
IP 54
Fastening and space required
Ø8h6 (3/8") fastening, interchangeable M2.5 (4-48-UNF) measuring probe (to DIN
878)
13.2.
BLUETOOTH
Data
Value
Frequency band
2.4 GHz
Bluetooth version
5.0
Modulation method
GFSK (Gaussian frequency shift keying)
Maximum transmission output capacity
Class 3: 1 mW (0 dBm)
Range
Outdoors: up to 45 m
In industrial environments: 1 – 5 m
28