Grounds and Floating Operations
47
If the
Counter
and
Working
fuses open when you hook up your cell, you probably have an earth ground in
your cell system and an earth-grounded test system. Disconnecting the jumper between
System Ground
and
Protective Ground
is likely to remedy this problem.
If the problem reoccurs with the ground jumper disconnected, make sure you do not have additional test
equipment connected to the system. If it still occurs with all additional equipment disconnected, contact
Gamry Instruments’ technical support staff.
Figure 5-3 shows the ground jumper in place.
Figure 5-3
Jumper Connection between Protective Ground and System Ground
When earth-grounding the System Ground is not desired, rotate the ground jumper out of the way but left
attached to one of the binding posts (Fig. 7-2). This will prevent it being lost. If the jumper has been misplaced,
you can run a wire between the binding posts to earth-ground the system.
Warning:
Connection of an earth-grounded cell to an earth-grounded Reference 30k
Booster can short the cell. This can create excessive current and hazardous conditions. The 30 A
Counter
and
Working
fuses provide only limited protection.