Installation – Optional 30 A Test
30
Figure 3-3
Connections to the Test Side of the Test & Cal Cell
Apply currents by selecting
Experiment > Utilities > Set a Current
on the Framework menu. In the window
that appears, enter a current density of 3.0E4 mA/cm
2
and an area of 1.0 cm
2
. After you click the
OK
button, a
runner window displays a current and voltage versus time curve that looks similar to Figure 3-4.
Figure 3-4
Current-versus-Time Curve for 30 A Applied to the Test Cell
The voltage on the plot should be between 88 and 92 mV. The current should be between 29 and 31 A.
As a further check, you can repeat this test applying –30 A. The nominal voltage and current values will be –
90 mV and –30 A.