![Gamry Instruments Reference 30k Booster Operator'S Manual Download Page 43](http://html1.mh-extra.com/html/gamry-instruments/reference-30k-booster/reference-30k-booster_operators-manual_3396599043.webp)
Reference 30k Booster Specifications
43
Appendix A:
Reference 30k Booster Specifications
All specifications are at an ambient temperature of 25
C, 115 V
AC
power, and the Reference 30k Booster
operated with a Reference 3000 potentiostat unless otherwise noted.
All specifications are subject to change without notice.
Control Amplifier
Compliance Voltage
+20 to –2.5
V
Compliance Current
±30
A
Voltage Gain
2
Unity-gain Bandwidth
500
kHz
Slew Rate
>20
V/µs
30 A Current Range
Full-scale Current
±30
A
Accuracy
±0.3
% of range
System (all with a Reference 3000 potentiostat)
Control Modes
Potentiostatic, Galvanostatic, ZRA
Allowed Current Ranges
30, 3, 0.3, 0.03
A
Mechanical
Dimensions (w x h x d)
37 x 23 x 44
14.5 x 9 x 17.5
cm
inch
Weight
16
35
kg
lbs
Power
Input Voltage
90 to 263
V
AC
Power (max.)
1200
W
A potentiostat system with a Reference 30k Booster connected to a Reference 3000 Potentiostat will not
oscillate with any capacitive load as long as the Reference electrode impedance is less than 10 kΩ and the
control amplifier bandwidth is Normal or slower.
A galvanostat system with a Reference 30k Booster connected to a Reference 3000 Potentiostat will not
oscillate with any resistive load less than 100
Ω
as long as the control amplifier bandwidth is Normal or slower.
This is only guaranteed when the cell is connected using Gamry’s P/N 985-00122 cable. Stability with inductive
loads is not guaranteed.