PortHole™ Electrochemical Sample Masks

 

PortHole

 Electrochemical Sample Masks 

PortHole Electrochemical Sample Masks define a specified area 
of a flat metal sample for electrochemical testing with the PTC1. 
PortHole masks are available in areas of 1, 3, or 10 cm

2

 to 

accommodate any specimen. PortHole masks are fabricated 
from vinyl 

3M™ Electroplating Tape 470

 

 

 

Before applying the PortHole mask, degrease the surface 
using a suitable solvent. Remove the backing from the 
PortHole mask and place it carefully on the metal sample. 
Roll the PortHole mask over the metal sample to make 
sure that the tape does not crimp around the perimeter of 
the hole. 

Carefully press down on the entire surface of the 
PortHole mask. Pay special attention to the perimeter of 
the hole: it is important to get good adhesion on the 
perimeter to minimize the possibility of crevice corrosion. 
Use a spatula or a round implement to press the edge of 
the tape firmly against the metal sample. 

Place the cell body and O-ring of the PTC1 on the PortHole mask so that the exposed area of the sample is 
enclosed. Clamp the cell body firmly as described in Cell Setup, and proceed normally. 

When the experiment is complete, remove the PortHole mask and discard. If the sample-electrolyte system 
tends to undergo crevice corrosion, visually examine the periphery of the hole with a magnifier. If any crevice 
corrosion is detected, discard the data for that sample. 

If crevicing is particularly prevalent with any sample, consider the following approach. Place the sample in a 
benign electrolyte and apply +0.75 volts for 60 seconds to form an oxide film. Remove the sample from the 
electrolyte and dry it. Place a PortHole mask in position on the sample. Manually remove the oxide film from 
the area exposed by the PortHole mask using a fine emery cloth or a pencil eraser. Perform the electrochemical 
experiment. 

 

 

PortHole™ masks.

 

Summary of Contents for PTC1

Page 1: ...PTC1 Paint Test Cell Kit Operator s Manual Copyright 2017 Gamry Instruments Inc Revision 2 1 September 15 2017 900 00057...

Page 2: ......

Page 3: ...obligation to install same on systems previously purchased All product specifications are subject to change without notice There are no warranties which extend beyond the description herein This warr...

Page 4: ...astern Standard Time 877 367 4267 Toll free US Canada Only Replacement parts for this kit are available from Gamry Instruments Inc Optional parts may be available from Gamry Instruments or from third...

Page 5: ...Damage 5 Parts List 5 Assembly of the Cell Top 6 Important notes 6 Procedure 6 Requirements for Metal Samples 7 Chemical Compatibility 7 Cell Setup 8 Working Electrode Area 8 PortHole Electrochemical...

Page 6: ...ing This cell was originally developed by Dr Richard Granata of Lehigh University The PTC1 was designed for electrochemical impedance testing of flat coated metal specimens It can also be used for tes...

Page 7: ...Gamry should have replacement parts in stock Please retain the shipment s packaging material for a possible claim against the shipping company Parts List Please check the contents of your kit versus t...

Page 8: ...rs 2 Gently push the graphite counter electrode into the smaller of the two holes in the top 3 Insert the counter electrode until approximately 5 cm of the electrode extends below the top 4 Wet the ou...

Page 9: ...quire testing without the base Painted samples must have an area outside of the cell free from paint This area is used to make electrical contact with the sample Chemical Compatibility The O ring used...

Page 10: ...op with the reference and counter electrode already in place on the cell Make sure that the protective cap is removed from the reference electrode With the top lightly pressed into the cell body you s...

Page 11: ...ity of crevice corrosion Use a spatula or a round implement to press the edge of the tape firmly against the metal sample Place the cell body and O ring of the PTC1 on the PortHole mask so that the ex...

Page 12: ...not in use Use the small protective cap shipped with the reference to cover the electrode tip if you want to store the top dry Many chemists believe that reference electrodes should be stored immerse...

Page 13: ...8 damage 5 distilled water 10 electrode area iii 8 Electroplating Tape 470 9 KCl 10 maintenance iii 10 O ring 5 7 8 9 10 Parts List 5 PortHole iii 9 Problems ii reference electrode iii 6 8 10 sample m...

Page 14: ......

Page 15: ......

Page 16: ...734 Louis Drive Warminster PA 18974 Tel 215 682 9330 Fax 215 682 9331 www gamry com info gamry com...

Reviews: