Gamry Instruments PTC1 Operator'S Manual Download Page 10

Cell Setup 

Cell Setup 

1.

 

Test your thin samples in a setup that includes the PTC1 base (a polypropylene block equipped with 
four rubber feet). Place a sample on the side of the base without the feet. 

2.

 

Place the cell body on top of the sample, making sure that the O-ring fits in the groove on the cell 
body. This O-ring makes a watertight seal between the cell body and the sample. 

3.

 

With the clamp, affix the glass to the base. Use 
the knurled knob on the cell clamp to 
compress the O-ring. 

 

 

 

 

 

 

4.

 

Fill the cell with the test solution (approximately 40 mL of solution). 

5.

 

Place the top, with the reference and counter electrode already in place, on the cell. Make sure that 
the protective cap is removed from the reference electrode. With the top lightly pressed into the cell 
body, you should have a watertight seal. 

6.

 

Before making the connections to the test instrument, check that the cell is not leaking from the O-ring 
seal. After you are sure the cell is not leaking, connect the test instrument. The metal sample is used as 
the working electrode.  

Working Electrode Area 

The area of the sample exposed to the test solution varies somewhat because of variation in the compression in 
the O-ring and variation in the glass cell body. Gamry Instruments measured the active area to be nominally 
14.6 cm

2

. At worst, the actual area should be within ±5% of this value. 

 

 

Summary of Contents for PTC1

Page 1: ...PTC1 Paint Test Cell Kit Operator s Manual Copyright 2017 Gamry Instruments Inc Revision 2 1 September 15 2017 900 00057...

Page 2: ......

Page 3: ...obligation to install same on systems previously purchased All product specifications are subject to change without notice There are no warranties which extend beyond the description herein This warr...

Page 4: ...astern Standard Time 877 367 4267 Toll free US Canada Only Replacement parts for this kit are available from Gamry Instruments Inc Optional parts may be available from Gamry Instruments or from third...

Page 5: ...Damage 5 Parts List 5 Assembly of the Cell Top 6 Important notes 6 Procedure 6 Requirements for Metal Samples 7 Chemical Compatibility 7 Cell Setup 8 Working Electrode Area 8 PortHole Electrochemical...

Page 6: ...ing This cell was originally developed by Dr Richard Granata of Lehigh University The PTC1 was designed for electrochemical impedance testing of flat coated metal specimens It can also be used for tes...

Page 7: ...Gamry should have replacement parts in stock Please retain the shipment s packaging material for a possible claim against the shipping company Parts List Please check the contents of your kit versus t...

Page 8: ...rs 2 Gently push the graphite counter electrode into the smaller of the two holes in the top 3 Insert the counter electrode until approximately 5 cm of the electrode extends below the top 4 Wet the ou...

Page 9: ...quire testing without the base Painted samples must have an area outside of the cell free from paint This area is used to make electrical contact with the sample Chemical Compatibility The O ring used...

Page 10: ...op with the reference and counter electrode already in place on the cell Make sure that the protective cap is removed from the reference electrode With the top lightly pressed into the cell body you s...

Page 11: ...ity of crevice corrosion Use a spatula or a round implement to press the edge of the tape firmly against the metal sample Place the cell body and O ring of the PTC1 on the PortHole mask so that the ex...

Page 12: ...not in use Use the small protective cap shipped with the reference to cover the electrode tip if you want to store the top dry Many chemists believe that reference electrodes should be stored immerse...

Page 13: ...8 damage 5 distilled water 10 electrode area iii 8 Electroplating Tape 470 9 KCl 10 maintenance iii 10 O ring 5 7 8 9 10 Parts List 5 PortHole iii 9 Problems ii reference electrode iii 6 8 10 sample m...

Page 14: ......

Page 15: ......

Page 16: ...734 Louis Drive Warminster PA 18974 Tel 215 682 9330 Fax 215 682 9331 www gamry com info gamry com...

Reviews: