2. GettInG ready for usaGe
2.1. assemblInG the devIce
1. Take the unit out of the box.
. Open the jet nebulizer (Fig. A) and pour the medica-
tion to nebulize into it.
Do not pour more than 10 ml (Fig. B).
Then twist both parts and close again. (Fig. C).
. Connect the tubing to the jet nebulizer and the
other end to the nebulizer outlet. (Fig. D).
. Put the mask or the mouth piece to be used into the
jet nebulizer (Fig. E).
. Connect the unit to a suitable power receptacle.
Your nebulizer system is ready for work.
2.2. replacement of the fIlter
1. Filter should be changed every month or sooner if the filter turns completely
gray in color.
. Remove filter cap by grasping it firmly and pulling out the unit.
. Change the dirty filter of the cap with small, pointed object.
. Insert a clean San Up filter. Extra filters are available.
. Push the filter cap with new filter into position.
caution:
Do not use cotton or any other material. Do not wash or clean the filter. Use only original filters. Do
not operate without a filter.
3. carryInG out aerosol therapy
1. Turn the unit by pressing the ON/OFF switch to start the treatment.
attentIon:
This device is for intermittent use: 0 min. ON/0 min. OFF.
. Inhale the aerosol solution using the prescribed accessory in calm way. One should sit at the
table with one’s back upright, relaxed; otherwise you will constrict your respiratory passages and
impair the effect of inhalation.
. Turn the unit off and unplug it once the treatment has finished.
. Clean the jet nebulizer and the devices as indicated in Cleaning and disinfection section.
Store the jet nebulizer and the devices into the unit’s lodging.
4. cleanInG and dIsInfectIon
Important:
1. Wash all the pieces, except for the air filter, with warm water and detergent. Rinse with warm tap
water to remove the detergent.
. Dip the pieces into a clean bowl containing parts of warm water and 1 part of vinegar during
0 minutes, disinfectant can also be used.
. Wash your hands and take out the pieces from the disinfectant solution, rinse them with warm
tap water and leave them on an absorbent piece of paper.
. Keep the pieces into a hermetic bag. Do not dry them with a towel.
. Clean the outer surface with a dry piece of cloth.
• Do not use any volatile fluid like alcohol or benzene.
• The jet nebulizer and the devices must NOT be boiled.
• The jet nebulizer has to be replaced if it is stuck by medication or other substances.
• Under no circumstances will the unit be opened.