Hex
command
Description
Data field Response
Data/Response description
Sets the calibration
factor that modifies
pressure.
33
SET AUTO-RESTART
Sets supply to
automatically
restart on power
up.
"YES" or
"NO"
34
GET AUTO RESTART
Reads auto restart
status of supplies.
"YES" or
"NO"
37
START PUMP
Enables high
voltage.
None or
"1"
38
STOP PUMP
Disables high
voltage.
None or
"1"
3C
GET SETPOINT
Reads configuration
of the set point.
None or
"1"
N, E, X.XE-XX,
Y.YE-YY, O
N = Setpoint Number
E = Enabled (1 = yes, 0 = no)
X.XE-XX = On Pressure
Y.YE-YY = Off Pressure
O = Setpoint On (1 = yes, 0 =
no)
3D
SET SETPOINT
Configures a
specified set point.
N, S,
X.XE-
XX,
Y.YE-
YY
N = Setpoint Number
S = Supply (0 = Inactive, 1 =
Supply 1) X.XE-XX = On
Pressure
Y.YE-YY = Off Pressure
44
LOCK KEYPAD
Established Remote
Mode and locks the
front panel except
HV off and power
keys.
45
UNLOCK KEYPAD
Unlocks all front
panel keys.
04/2021 - ©Gamma
Page 45
900026_F
900026_F - Serial (J3) and ethernet (J2) operation