WRO 300 H
HCEN12481-02/2012
Troubleshooting
69
5.1.6
Perform residual test
5.1.7
Reminders
No.
Display text
Alarm
type
Definition
“None”
PERFORM
RESIDUAL
TEST
WRO 300 H
The WRO 300 H unit stops, low priority
alarm is issued,
flashes, message
text is displayed. The indication
will
remain steady lit and the
will flash.
Alarm sys-
tem delay
None
Reason
A chemical disinfection/cleaning/preser-
vation program is completed and a
residual test is required.
Action
Perform a residual test according to
section 3.
No.
Display text
Alarm
types
Definition
“None”
“User defined
text”
WRO 300 H
When entering Standby mode, low prio-
rity alarm is issued,
flashes,
message text is displayed.
Alarm sys-
tem delay
None
Reason
A user definition reminder has been set
to remind the user to take certain actions.
Action
1
Press
to silence buzzer.
2
Perform action according to the
reminder.