WRO 300 H
HCEN12481-02/2012
Troubleshooting
61
203
203 CONDUC-
TIVITY STOP
WRO 300 H
WRO 300 H stops, low priority alarm is
issued,
flashes,
message text is displayed.
Alarm sys-
tem delay
30 seconds
Reason
The conductivity value has exceeded the
stop limit for more than 30 consecutive sec-
onds.
Action
1
Press
to silence buzzer
2
Inform responsible medical staff accord-
ing to local procedures
3
The WRO 300 H can be restarted by
pressing
. It will stop again if the
problem persist.
4
If problem persists discontinue dialysis
5
Notify technical personnel.
204
204 INSUFFI-
CIENT WATER
SUPPLY
WRO 300 H
WRO 300 H stops, low priority alarm is
issued,
flashes, message text is
displayed.
Alarm sys-
tem delay
None
Reason
Lack of feed water.
Action
1
Press
to silence buzzer
2
Check the feed water supply
3
Notify technical personnel.
No.
Display text
Alarm
type
Definition