WRO 300 H
58
Troubleshooting
HCEN12481-02/2012
5.1.2
General
No.
Display text
Alarm
type
Definition
101
101 CONDUC-
TIVITY SEN-
SOR FAILURE
WRO 300 H
Low priority alarm is issued,
flashes,
message text is displayed.
Alarm sys-
tem delay
5 minutes
Reason
The initial conductivity system check failed.
Action
1
Press
to silence buzzer
2
Check that the product water conductivity
is within normal range. If so, continue as
normal but notify technical personnel.
102
102 HIGH
LEVEL SEN-
SOR FAILURE
WRO 300 H
WRO 300 H stops, low priority alarm is
issued,
flashes,
message text is displayed.
Alarm sys-
tem delay
None
Reason
Control unit failure.
Action
1
Press
to silence buzzer
2
Restart unit, or notify technical
personnel.
103
: N INTERNAL
ERROR
a
or
WRO 300 H
WRO 300 H stops, low priority alarm or infor-
mation signal is issued,
flashes,
message text is displayed.
Alarm sys-
tem delay
Between 5 seconds and 60 minutes.
Reason
Control unit failure.
Action
1
Press
to silence buzzer
2
Restart unit, or notify technical
personnel.