LED Program Board Overview
COBE 2991 Cell Processor
•
Essentials Guide
4-5
LED Progra
mming
4
VALVE SELECT
VALVE SELECT determines which valve opens during the Agitate/Wash-In
operation.
Select or deselect the desired valve by pressing the key below the appropriate
VALVE SELECT LED. Only one valve can be programmed to open during the
Agitate/Wash-In operation. A different valve can be selected while the
Agitate/Wash-In operation is in progress by pressing the appropriate key. If no
valves are programmed to open, the COBE 2991 goes into a Hold status at the
beginning of the Agitate/Wash-In operation.
For more information on programming no valves, see “Programming a Hold at the
Beginning of the Agitate/Wash-In Operation” on page 4-6.
Programming a Hold
In addition to using the CALL stop option to program a Hold status at the
beginning of the Supernatant-Out operation, you can also program a Hold status at
the beginning of the Start/Spin or Agitate/Wash-In operations using the following
methods.
Programming a Hold Status at the Beginning of the Start/Spin Operation
To program a Hold status at the beginning of a Start/Spin operation for a given
cycle, program a zero spin time for that cycle.
When the Start/Spin operation with the zero spin time programmed is reached, the
HOLD light illuminates, the centrifuge continues spinning, the primary audible
alarm sounds (if turned on), and the secondary audible alarm sounds.
You have the following options:
•
Enter a time of 0.1 minute or greater and press CONTINUE. The Start/Spin
operation continues for the time entered.
-or-
•
Press the STOP RESET button to end the program.
To display the total time elapsed since the COBE 2991 entered the Start/Spin
operation, press the ELAPSED TIME key. The time displayed includes the time
spent in Hold status and any time that was entered at the end of the Hold status.
Summary of Contents for COBE 2991
Page 1: ......
Page 18: ...This page intentionally left blank ...
Page 19: ...This page intentionally left blank ...
Page 20: ...This page intentionally left blank ...
Page 22: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ...Hướng dẫn sử dụng Part No 770273 010 Reorder No 709016 000 2005 05 Máy xử lý tế bào ...
Page 27: ...COBE 2991 Cell Processor Essentials Guide 1 1 Introduction 1 1 Giới thiệu ...
Page 48: ...LED Programming 4 COBE 2991 Cell Processor Essentials Guide 4 1 4 Lên chương trình LED ...
Page 105: ...Troubleshooting 7 COBE 2991 Cell Processor Essentials Guide 7 1 7 Xử lý sự cố ...
Page 116: ...General Maintenance 8 COBE 2991 Cell Processor Essentials Guide 8 1 8 Bảo dưỡng ...
Page 120: ...Essentials Guide 1 6 17 05 10 29 08 AM ...
Page 121: ...Essentials Guide Part No 770273 010 Reorder No 709016 000 2005 05 Cell Processor ...