Leaking on solenoid
valve
1.
Solenoid valve obstructed. Attention: It is of high
importance to check the machine after installation as lime
scale or debris may be able to fall into the valve during
installation. Shut off the water supply and take out the valve.
Clean the valve and put it back to its position.
Continue fill or cannot
fill water
1.
Tiny obstruction entered the fill solenoid valve, resulting
in abnormal functioning. Caution: detail checking right
after installation is very important, small chips may go
inside into the piping and being stuck inside valve. Power
off, open the solenoid valve and clean all the interior
parts, then fix up.
2.
Inlet water flow too low.