background image

5

LI0196  –  19 July 2017

Principles of Operation

The 42AG consists of a sound source – a small loudspeaker – a reference microphone, and an 
electronic control circuit. The special reference microphone is extremely stable and independent of 
variations in barometric pressure, temperature, and humidity. Together, the reference microphone 
and control circuit maintain a constant sound pressure level inside the calibration coupler, making 
it possible to calibrate without separately accounting for the ambient calibration conditions and 
changes to the load volume.

42AG can produce a sinusoidal signal of 250 Hz or 1 kHz at 94 dB or 114 dB. The environmen-
tal conditions: air pressure, temperature and humidity can be displayed.

The calibrator complies with IEC 60942 (2003), Class 1 and has been designed to serve 1" and 
smaller microphones and sound level meters equipped with such microphones.

42AG is designed for use in the field. The accuracy that can be achieved is within ±0.2 dB which 
is more than adequate for field checks. 

For proper laboratory calibration, we recommend the use of a pistonphone, or sending the micro-
phone to a calibration laboratory.

Sinus generator

Amplifier

Feed-back and 

Control Circuit

Coupler

Microphone 

to be 

calibrated

Venting

Sinus Generator

Amplifier

Feed-back and

Control Circuit

Coupler

Microphone 

to be 

calibrated

Venting

Fig. 1. 

Simplified block diagram illustrating 42AG's principles of operation.

Caution. 

The loudspeaker inside the calibration coupler is not protected. Therefore, do not touch  

the diaphragm and also take care to prevent any hard objects from entering the coupler as this 
may deform or damage the diaphragm. 

Summary of Contents for 42AG

Page 1: ...Instruction Manual www gras dk LI0196 19 July 2017 G R A S 42AG Multifunction Sound Calibrator ...

Page 2: ...tp www gras dk Any technical documentation that is made available by G R A S is the copyrighted work of G R A S and is owned by G R A S The content in this document is subject to change without notice G R A S Sound Vibration A S is not liable or responsible for any errors or inaccuracies that may appear in this document Trademarks Any product names mentioned in this document may be trademarks or r...

Page 3: ... 6 Using the 42AG 7 Microphone Calibration 7 Calibration Considerations 9 Accuracy 9 Background Noise 9 Corrections 9 Maintenance and Calibration 11 Replacing Batteries 11 Calibrating the 42AG Sound Calibrator 11 Adjusting the 42AG s Levels 11 Delivered Items 13 Technical Specifications 13 Warranty Service and Repair 14 ...

Page 4: ...uded in the delivery Green LED lights when calibration level is OK Automatic compensation for ambient air pressure temperature and humidity Measurement of ambient air pressure temperature in C or F and humidity Automatic power off after 10 to 20 seconds Automatic shut off when batteries are too low About this Manual This manual contains the main sections listed below When referring to 42AG s butto...

Page 5: ...ibrator complies with IEC 60942 2003 Class 1 and has been designed to serve 1 and smaller microphones and sound level meters equipped with such microphones 42AG is designed for use in the field The accuracy that can be achieved is within 0 2 dB which is more than adequate for field checks For proper laboratory calibration we recommend the use of a pistonphone or sending the micro phone to a calibr...

Page 6: ... during power on Hz Toggles between 1 kHz and 250 Hz Default is 1 kHz dB Toggles between 94 dB and 114 dB Default is 94 dB Status Indicator Lights green when the pressure in the coupler has stabilized Display Displays the actual values for calibration and ambient conditions o N l a i r e S 888888888 G R A S 42AG Multifunction Sound Calibrator ANSI ASA S1 40 2006 R2011 Class1 IEC 60942 Batteries 2x...

Page 7: ... frequency is indicated in the display of the calibrator After powering the sound calibrator the first combination of frequency level is 94 dB re 20µPa and 1 kHz Select the appropriate level and frequency for your calibration by pressing dB or Hz respectively Normally the background level should be more than 20 dB lower than the level from the calibrator in order to not influence the reading There...

Page 8: ...output voltage is Vo and the sound pressure level in the Pistonphone s coupler is LC decibels the microphone sensitivity S is given by Vo 20 μPa 10 Lc 20 S The value 20 μPa is the standard reference sound pressure The result here includes the load ing effect of the preamplifier s input impedance as well as the gain or attenuation within the preamplifier To measure the Open Circuit Sensitivity of t...

Page 9: ... to the nominal level Nevertheless it is considered good practice to note the ambient calibration conditions as part of each calibration routine Temperature The temperature may be displayed in either degree Celsius C or degree Fahrenheit F ln order to change the unit keep the button Met depressed during power on Load Volume Various types and makes of microphones may have different effective load v...

Page 10: ...ree field microphone is designed to compensate for the pressure build up that takes place when it is used in a free field When calibrated with a sound calibrator at 1 kHz the output from the microphone will be lower compared to the output at free field conditions because the calibrator performs a pressure calibration of the microphone Therefore the result of a calibration of a free field microphon...

Page 11: ...ivities at the different frequencies are known with a sufficiently high accuracy Adjusting the 42AG s Levels Caution It is possible to adjust 42AG s levels the SPL the displayed temperature air pressure and relative humidity However if you do this you will invalidate the factory calibration and break the link to previous measurements For this reason we have sealed the access to this function which...

Page 12: ...ashing 3 While the display is flashing press Hz to increase the reading or dB to reduce the reading 4 When the correct value is displayed press Met and the value for air pressure starts to flash 5 While the display is flashing press Hz to increase the reading or dB to reduce the reading 6 When the correct value is displayed press Met and the value for humidity start to flash 7 While the display is...

Page 13: ...evel 94 dB 0 2 dB 114 dB 0 2 dB Distortion 2 0 Sensitivity to environmental conditions IEC 60942 Class 1 Display of temperature air pressure and humidity Temperature 10 C to 50 C accuracy 2 C resolution 0 1 C Atmospheric pressure 65 kPa to 108 kPa accuracy 0 4 kPa resolution 0 1 kPa Relative humidity 25 to 90 accuracy 6 resolution 1 Microphone sizes 1 1 2 1 4 1 8 Power Two 1 5V LR03 AAA Size alkal...

Page 14: ...are damaged due to negligent use an incorrect power supply or an incorrect connection to the equipment Service and Repairs All repairs are made at G R A S International Service Center located in Denmark Our Service Center is equipped with the newest test equipment and staffed with dedicated and highly skilled engineers Upon request we make cost estimates based on fixed repair categories If a produ...

Reviews: