background image

 

D1020 

- SIL 2 Powered Isolating Driver Smart-Hart compatible 

G.M. International ISM0012-14 

Configuration 

Screwdriver 

for Trimmers 

Ch1 and Ch2 

1 x 0.5 mm 

Calibration Trimmer Ch1 

Span Trimmer 

CH1 

Calibration Trimmer Ch2 

Side B Panel View 

Turn the trimmer clockwise to 
Increase Output value or turn 

the trimmer counterclockwise 

if you want to decrease Output 

value. 

Side A Panel View 

Span Trimmer 

CH2 

 

Calibration instruction and Troubleshooting 

In case a calibration check or readjustment is necessary the following equipment is required: 



Set power supply (PS) at 24.0 V at terminals 3 (+) and 4 (-), the calibrator (CAL) in the function “mA” 20 mA range at terminals 1 (+) and 2 (-), the multimeter (V1) for the reading in 
DC supply 30 V range at terminals 1 (+) and 2 (-), the multimeter (DMM) for the reading in DC supply 20 mA range at terminals 14 (+) and 15 (-). 



Execute the next test to verify D1020S unit channel or execute it twice to verified both D1020D unit channels. 



Verify that green led on the front panel is ON. 



Connect the unit as shown below: 

If the power ON Led is off, check supply voltage, polarity and wiring. 



Set calibrator (CAL) at 4.000 mA and verify that the output current, read by multimeter (DMM), is correspondent with a maximum deviation presented in the section “Technical Data”. 



Set calibrator (CAL) at 20.000 mA and verify that the output current, read by multimeter (DMM), is correspondent with a maximum deviation presented in the section “Technical Data”. 



Set calibrator (CAL) at 20.000 mA and power supply (PS) at 20.0 V; verify that multimeter (V1) reads input voltage as presented in the section “Technical Data”. 

If the maximum deviation is exceeded, repeat the board calibration and proceed in the following way:  



Set the calibrator (CAL) at 20.000 mA and regulate the current, read by multimeter (DMM), by turning the span trimmer of the unit (see “Configuration” section) until reading 20.000 
mA ± 5 µA value. 



To execute a linearity test: set the calibrator (CAL) at 4.000, 8.000, 12.000, 16.000, 20.000 mA and verify current output is correspondent with 
a maximum deviation presented in the section “Technical Data”. 

Adjustable Power Supply Range 0-35 V Resolution 0.1 V, 
Accuracy 1% Output Capability 200 mA 
Current Meter range 20 mA, Resolution 1 µA Accuracy 0.01% 
or Voltage Meter Range 10 V, Resolution 1 mV Accuracy 0.01% 

Voltage Meter range 30 V, Resolution 1 mV Accuracy 0.05 % 

Current Calibrator range 20 mA, Resolution 1 µA Accuracy 0.01 % 

Agilent 

E3611A 

Agilent 

34410A 

Fluke 

175 

Yokogawa 

7651 

Equipment Required 

Instrument 

PS 

DMM 

V1 

CAL 

Identification 

PS

Supply

=

_

+

V1

CAL

DMM

mA

mA

MODEL D1020D

=

=

=

14

15

3 +

4 -

1

2

=

=

=

10

11

5

6

=

=

=

=

In 1

V1

CAL mA

Out 1

+

DMM

mA

Out 2

7 +

8 -

In 2

Summary of Contents for D1020S

Page 1: ...D1020 SIL 2 Powered Isolating Driver Smart Hart compatible ISM0012 14 D1020S D1020D INSTRUCTION SAFETY MANUAL SIL 2 Powered Isolating Driver Smart Hart compatible DIN Rail Models D1020S D1020D...

Page 2: ...ompatibility CE mark compliant conforms to Directive 2014 34 EU ATEX 2014 30 EU EMC 2014 35 EU LVD 2011 65 EU RoHS Environmental conditions Operating temperature limits 20 to 60 C relative humidity ma...

Page 3: ...tions High Reliability SMD components High Density two channels per unit Simplified installation using standard DIN Rail and plug in terminal blocks 250 Vrms Um max voltage allowed to the instruments...

Page 4: ...cable and shall not be exceeded 50 of the Co and Lo become the limits which must include the cable such that Ci device C cable 50 of Co and Li device L cable 50 of Lo If the cable parameters are unkno...

Page 5: ...G CLASS III DIVISION 1 CLASS I ZONE 0 GROUP IIC SAFE AREA ZONE 2 GROUP IIC T4 NON HAZARDOUS LOCATIONS CLASS I DIVISION 2 GROUPS A B C D T Code T4 CLASS I ZONE 2 GROUP IIC T4 MODEL D1020S 14 15 3 4 1...

Page 6: ...afe state or deviates the output current by more than 5 full span 0 8 mA Fail High failure mode that causes the output signal to go above the maximum output current 20 mA This failure mode is consider...

Page 7: ...Division 2 or Class I Zone 2 location the wiring between the control equipment and the D1020 associated apparatus shall be accomplished via conduit connections or another acceptable Division 2 Zone 2...

Page 8: ...ive and 15 for negative For Model D1020D in addition to channel 1 connections above connect terminal 10 for positive and 11 for negative on channel 2 Intrinsically Safe conductors must be identified a...

Page 9: ...deviation presented in the section Technical Data Set calibrator CAL at 20 000 mA and verify that the output current read by multimeter DMM is correspondent with a maximum deviation presented in the...

Reviews: