background image

 

 

When the T3 device is connected to the computer the device display lights up and shows a 
message indicating that the device is connected to the computer. 
 
To configure a T3 device or to download a recording you will now need to start the KOEO Sleep 
SMART Process.  
 
When you are done working with the T3 device, unplug the mini USB cable, insert the battery, 
and close the battery compartment by 

gently pressing

 the lid towards the device, and then 

sliding it back into position, towards the top of the device. 
 

Inserting a Battery into the T3 Device 

Before you start a recording you should make sure that the device has a new, fully charged AA 
battery. To insert a new battery do the following: 
 

 

1. Open the battery compartment by pressing down on the battery lid pin with the tool 

provided, a pen, or similar tool and then slide the lid toward the bottom of the device (left 
image). 

2. Place one AA battery into the compartment aligning the battery poles as illustrated on 

the back of the device (the positive (+) pole is toward the battery lid). 

3. Close the battery compartment by 

gently pressing

 the lid toward the device while sliding 

it up and back into position towards the top of the device. Make sure the lid is securely 
closed. 

 

Summary of Contents for Sleep SMART Nox T3

Page 1: ...T3 device to a computer you will need to access the USB connector on the device The USB connector is located under the battery lid making it inaccessible and tamper proof for children To open the batt...

Page 2: ...rting a Battery into the T3 Device Before you start a recording you should make sure that the device has a new fully charged AA battery To insert a new battery do the following 1 Open the battery comp...

Page 3: ...w Subject Do not enter the subject s name or Date of Birth Instead enter the following fields First Sleep SMART Box 1 Last Subject ID Number Box 2 Date of Birth 1 1 1900 Box 3 Identification Subject I...

Page 4: ...ading testing data and must be already installed on the computer you use for the set up The two main functions of KIT are as follows Set up a T3 device Upload test data from a T3 device KIT operations...

Page 5: ...e must be connected via a mini USB cable to the computer at this point The next step configures the T3 Device Step 5 T3 Setup Task in KIT Select the appropriate Time Zone from the drop down menu for t...

Page 6: ...l now proceed to Upload T3 Data Step 6 This task is to be completed only after the T3 test is complete Upon completion of the T3 test in Sleep SMART the data must be uploaded to the scoring teams for...

Page 7: ...xt Tasks in KOEO The Sleep SMART Protocol has provisions for repeated attempts under appropriate circumstances to collect data from the subject using the Nox T3 device A number of different task optio...

Page 8: ...re are extenuating circumstances that by protocol allow you to repeat the study you may do so by starting from page 1 instructions Connecting the T3 device to the computer If the Repeat T3 Setup task...

Page 9: ...con in the upper right hand corner to print out a physical copy of the Sleep SMART Sleep Apnea Test Report When printing is complete or if printing is not necessary e g subject is deceased click the c...