14
Calibration
1. Once the scale has its power ON and is in Weighing Mode, use a flathead
screwdriver to gently remove the Calibration Seal Screw located in the
rear of the scale.
Display in
Calibration Mode
Calibration of the DRX-4 should be performed with a single accurate 100g
weight. To Calibrate the scale follow the directions below:
2. Once the screw has been removed and the calibration switch exposed,
use a pen or another pointing device to toggle the calibration switch to the
right (which is the ON position.)
3. Upon toggling of the calibration switch, the balance will go into Calibration
Mode. For approximately five (5) seconds the display will read CAL,
indicating the scale is in Calibration Mode.
I n s t r u c t i o n M a n u a l M o d e l D R X - 4
C H A P T E R 1 1