background image

UNLOCKING THE BALANCE 

 

1.

 

Be sure the small operating knob in the lower front center of the base is turned 
clockwise to lock the balance. Lay the balance on its back and loosen the two 
knurled shipping arrest knobs (A) under the base (shown in figure 1). Do not 
touch the screw in the center of the knobs. Merely loosen the knobs slightly, push 
them as far as possible towards the ends of the balance, and tighten them again as 
shown in figure 2. 

2.

 

Place the balance upright on a firm, level vibration-free with the dial and 
operating knob facing the operator. 

3.

 

Place the pan holders on top of the trusses visible through the grommet openings 
in the glass on the top of the balance (Figure 2). Place the pans on the pan holders. 
If the balance has a scoop, it should be placed on the left-hand pan holder. 

 
 

LEVELING THE BALANCE 

 

This is accomplished by turning both leveling screw in the same direction at the 

same time and in equal amounts. This will raise or lower the balance as required to bring 
the bubble in the level, located on the lower right side of the case, to the center of the 
vial.  
 

ZEROING THE BALANCE 

 
Set the dial knob to zero. Release the balance by turning the operating knob in the lower, 
front center of the base counterclockwise. Bring the balance into equilibrium by means of 
the leveling screws as follows: If pointer moves to the right, turn the left leveling screw 
counterclockwise and the right leveling screw clockwise. The leveling screws should be 
rotated simultaneously and as equally as possible. If the pointer moves to the left, turn the 
left leveling screw clockwise and the right leveling screw counterclockwise. 
 
The balance is now ready for use. If it does not operate properly, recheck the directions to 
be certain that all steps have been performed. Do not attempt any adjustments you may 
think necessary, other than those indicated in the instructions, without first contacting the 
factory. When you are satisfied that the balance is in proper operating condition, please 
complete and mail the post paid warranty card. 
 

NOTICE 

 
Regulations regarding the inspection of prescription balances vary from one jurisdiction 
to another. To eliminate the possibility of any problems, you should, within 24 hours of 
receipt of your balance, contact your local Weights and Measures office to arrange for an 
inspection if it is required. 
 

 
 