ix
Precautions about Output Signal
WARNING
Do not use the output signal of the monitor that receives radio wave signal from the
LX-8100/LX-8100(G) as the trigger signal for IABP, MRI echocardiographic, or defibrillator for
the following reasons.
It may lead to a delay of operating timing due to the delay time of waveform transmission.
A trigger signal unrelated to the heart rate may be generated due to the interfusion of spike noise at
weak electric field.
Precautions about Accessories and Optional Accessories
WARNING
Use only the accessories such as disposable electrodes and electrode codes specified by
Fukuda Denshi. Otherwise, this equipment cannot deliver its maximum performance and may
be damaged, resulting in a safety hazard.
CAUTION
Do not reuse disposable products.
Store the disposable products properly as mentioned in their user manuals.
Precautions about Battery
WARNING
Use new "AA" size (“LR6” size) alkaline cell.
Install the battery with the correct polarity.
Do not charge the battery. Any attempt to charge the battery may cause it to leak or break.
Do not short the (+) and (-) terminals. It may result in exothermic heat and fire.
Do not throw the battery into fire. It may explode.
Precautions about Nickel Metal Hydride Battery (for Europe)
WARNING
Use "AA" size (“HR6” size) fully charged nickel metal hydride battery.
Install the battery with the correct polarity.
Read the manual of the battery, charger and accessories prior to use.
Do not short the (+) and (-) terminals. It may result in exothermic heat and fire.
Do not throw the battery into fire. It may explode.
Precautions about Disposing of Equipment, Accessories, or Components
CAUTION
When disposing of equipment, accessories, or components, use an industrial waste
distributor. Do not dispose of as ordinary waste.
Used disposal items (ECG electrodes, etc.) shall be discarded as medical waste.