1.  Use  a  medical  grade  mild  disinfectant  wipe  on  the  cuff  or  spray  a  cleaning 

solution onto a cloth and wipe the cuff. Afterwards, lay flat or line dry. 

2.  Remove the bladder and microphone from the outer shell of the Orbit-K cuff. 

Machine wash the shell in warm water with a mild detergent (50-140°F or 10-
60°C). Lay flat or line dry the cuff. Do not place the cuff in a dryer. 

 

Q. My Tango M2 displays a message, “Please VERIFY CALIBRATION” or 

“Equipment Maintenance and Calibration Required.” What do I do? 

A. 

Verification of Pressure Calibration is required every year to maintain the 
accuracy of the Tango M2’s BP readings. You will need to contact your local 
SunTech Medical Service Center for help. In addition, the following items will be 
needed to verify calibration. 

 

Equipment Required: 

1.  Calibrated electronic manometer or equivalent. 
2.  500mL volume or the Orbit-KAdult Plus cuff wrapped around something 

that will not break or crush (no glass). 

3.  Hand Inflation Bulb with bleed valve. 
4.  Tubing, Tee pieces, and miscellaneous connectors or you can order the 

T-Tube Kit (SunTech Part # 98-0030-00). 

 
 
 

Service Centers: 

For customers in the 
Americas SunTech Medical, 
Inc. 
507 Airport Boulevard 
Suite 117 
Morrisville, NC 
27560-8200 USA 
Phone: +1.919.654.2300 

+1.800.421.8626 (toll free 

within US) Fax: +1.919.654.2301 
Email: 

[email protected]

 

 
 
 
 

Summary of Contents for FX-8400

Page 1: ...rement and allows the Tango M2 BP reading to be transferred to the FX 8400 display and reports o SunTech Part Number 91 0010 01 or FUKUDA DENSHI Part Number CJS 03DA5 o Connection Ends FX 8400 Connection Tango M2 Connection Mini DIN 8pin male D sub 9 pin female l ECG Trigger Cable provides the ECG signal from the FX 8400 to the Tango M2 o SunTech Part Number 91 0018 02 or FUKUDA DENSHI Part Number...

Page 2: ...8400 3 Connect the ECG Trigger Cable From the BNC External ECG connection on the rear panel of the Tango M2 To the 1 8th phono plug connection on the side of the FX 8400 RS 232C Connection Part 91 0010 01 or CJS 03DA5 CJS 03DA5 ECG Triger Connection Part 91 0018 02 or CJS 02JB5 RS 232C Connection Part SunTech PN 91 0010 01 Or FUKUDA DENSHI PN CJS 03DA5 ...

Page 3: ...ECG Triger Connection PN 91 0018 02 or FUKUDA DENSHI PN CJS 02JB5 ...

Page 4: ...the Setting button c Touch the External Device button d Select BP Monitor Type to Tango e Touch the To exam button f When a confirmation message appears touch YES FAQs for Tango M2 Q The Tango M2 displays a Status Message What does it mean and what do I do A See the eLibrary within your Tango M2 for quick troubleshooting tips You can find the eLibrary under the Main Menu select Monitor Setup E Lib...

Page 5: ...brightness of the Tango M2 display A You can adjust the contrast of the Tango M2 display by following these steps 1 When the operating screen is displayed press the SELECT button once This will bring up the Main Menu screen 2 Using the UP or DOWN arrows highlight Monitor Setup and press the SELECT button 3 Using the UP or DOWN arrows highlight Brightness and press the SELECT button 4 Using the UP ...

Page 6: ... maintain the accuracy of the Tango M2 s BP readings You will need to contact your local SunTech Medical Service Center for help In addition the following items will be needed to verifycalibration Equipment Required 1 Calibrated electronic manometer or equivalent 2 500mL volume or the Orbit KAdult Plus cuff wrapped around something that will not break or crush no glass 3 Hand Inflation Bulb with b...

Page 7: ...urope the Middle East Africa Asia and the Pacific SunTech Medical Ltd Oakfield Industrial Estate Eynsham Oxfordshire OX29 4TS UK England Phone 44 0 1865 884 234 Fax 44 0 1865 884 235 Email CustomerSupport SunTechMed com ...

Reviews: