Look in the Tango M2 User Manual under the 

Using Tango M2 during a Stress 

Test 

section for details on each type of cuff; the Orbit-K and the Single Patient Use 
(SPU) kit. 
c.  Follow  the  instructions  in  the 

Cuff  Tutorial 

(located  on  the  SunTech  Medical 

website  under  Support  Customer  Technical  Support   Video  Tutorials)  for 
correct microphone placement. 

 

Q. Can I use a heart rate or blood pressure simulator to test whether the 

Tango M2 is working correctly with my stress system? 
A. 

You cannot use a heart rate or blood pressure simulator to test whether the 

Tango M2 is working with your stress system. The Tango M2 monitor requires 
that the ECG signal and the Korotkoff sounds, collected by the microphone in 
the cuff, originate from the same source, meaning the patient. 

 

Q. How can I adjust the brightness of the Tango M2 display? 
A. 

You can adjust the contrast of the Tango M2 display by following these steps: 
1.  When the operating screen is displayed, press the 

SELECT 

button once. 

This will bring up the 

Main Menu 

screen. 

2. 

Using the UP or DOWN arrows, highlight 

Monitor Setup 

and press the 

SELECT 

button. 
3.  Using the UP or DOWN arrows, highlight 

Brightness 

and press the 

SELECT 

button. 

4.  Using the UP or DOWN arrows, modify the contrast of the screen. 

When you are finished, and press the 

SELECT 

button to confirm the 

choice. 

5.  Using the UP or DOWN arrows, select 

EXIT 

twice to return to the operating 

screen. 

 

Q. How can I get MAP on my Tango M2? 
A. 

Register your Tango M2 Monitor online or through mail to receive MAP as a 

function on your Tango M2 monitor. Please note that due to FDA Regulations MAP 
is not available to the United States market. 

On the web go to SunTechMed.com  Support   Sales Support   Product 
Registration 

 

Q. How do I clean the Orbit-K cuff after a stress test? 
A. 

You can do either of the following: 

Summary of Contents for FX-8400

Page 1: ...rement and allows the Tango M2 BP reading to be transferred to the FX 8400 display and reports o SunTech Part Number 91 0010 01 or FUKUDA DENSHI Part Number CJS 03DA5 o Connection Ends FX 8400 Connection Tango M2 Connection Mini DIN 8pin male D sub 9 pin female l ECG Trigger Cable provides the ECG signal from the FX 8400 to the Tango M2 o SunTech Part Number 91 0018 02 or FUKUDA DENSHI Part Number...

Page 2: ...8400 3 Connect the ECG Trigger Cable From the BNC External ECG connection on the rear panel of the Tango M2 To the 1 8th phono plug connection on the side of the FX 8400 RS 232C Connection Part 91 0010 01 or CJS 03DA5 CJS 03DA5 ECG Triger Connection Part 91 0018 02 or CJS 02JB5 RS 232C Connection Part SunTech PN 91 0010 01 Or FUKUDA DENSHI PN CJS 03DA5 ...

Page 3: ...ECG Triger Connection PN 91 0018 02 or FUKUDA DENSHI PN CJS 02JB5 ...

Page 4: ...the Setting button c Touch the External Device button d Select BP Monitor Type to Tango e Touch the To exam button f When a confirmation message appears touch YES FAQs for Tango M2 Q The Tango M2 displays a Status Message What does it mean and what do I do A See the eLibrary within your Tango M2 for quick troubleshooting tips You can find the eLibrary under the Main Menu select Monitor Setup E Lib...

Page 5: ...brightness of the Tango M2 display A You can adjust the contrast of the Tango M2 display by following these steps 1 When the operating screen is displayed press the SELECT button once This will bring up the Main Menu screen 2 Using the UP or DOWN arrows highlight Monitor Setup and press the SELECT button 3 Using the UP or DOWN arrows highlight Brightness and press the SELECT button 4 Using the UP ...

Page 6: ... maintain the accuracy of the Tango M2 s BP readings You will need to contact your local SunTech Medical Service Center for help In addition the following items will be needed to verifycalibration Equipment Required 1 Calibrated electronic manometer or equivalent 2 500mL volume or the Orbit KAdult Plus cuff wrapped around something that will not break or crush no glass 3 Hand Inflation Bulb with b...

Page 7: ...urope the Middle East Africa Asia and the Pacific SunTech Medical Ltd Oakfield Industrial Estate Eynsham Oxfordshire OX29 4TS UK England Phone 44 0 1865 884 234 Fax 44 0 1865 884 235 Email CustomerSupport SunTechMed com ...

Reviews: