Fukuda Denshi FX-8400 Connection Manual Download Page 4

4.  Tango M2 Monitor Setup 

a.  When the operating screen is displayed, press the 

SELECT

 button once. This will bring up the 

MAIN MENU

 screen.

 

b.  Using the UP or DOWN arrows, highlight 

MONITOR SET UP

 and press the 

SELECT

 button.

 

c.  Using the UP or DOWN arrows, highlight 

STRESS SYSTEM

 and press the 

SELECT

 button.

 

d.  Using the UP or DOWN arrows, scroll through the list until 

CUSTOM

 is highlighted and press the 

SELECT

 button to confirm the choice.

 

e.  Using the UP or DOWN arrows, highlight 

PROTOCOL

 and press the 

SELECT

 button. 

f. 

Using the UP or DOWN arrows, highlight 

STANDARD

 and press the 

SELECT

 button. 

g.  Using the UP or DOWN arrows, highlight 

ECG TRIGER

 and press the 

SELECT

 button. 

h.  Using the UP or DOWN arrows, highlight 

DIGITAL FALLING

 and press the 

SELECT

 button. 

i. 

Using the UP or DOWN arrows, select 

EXIT

 twice to return to the operating screen. 

 

5.  FX-8400 System Setup 

a.  From the main screen, select 

MENU

 button.

 

b.  Touch the 

Setting

 button.

 

c.  Touch the 

External Device

 button. 

d.  Select 

BP Monitor Type

 to 

Tango+. 

e.  Touch the 

To exam

 button. 

f. 

When a confirmation message appears, touch 

YES.

 

 

FAQs for Tango M2 

 

Q. The Tango M2 displays a Status Message. What does it mean and what do I do? 
A. 

See the eLibrary within your Tango M2 for quick troubleshooting tips. You can   

find the eLibrary under the Main Menu; select Monitor Setup  E-Library  Status 
Messages. You can also look in the Tango M2 User Manual under the 

Status 

Messages & Alarms 

section for details on each Status Message and solution. 

 

Q. The Tango M2 monitor returns a result of 0/0 after blood 

pressure (BP) measurements. What do I need to do to get a 
BP reading? 

A. 

There are certain noisy conditions where the Tango M2 cannot accurately 
measure BP. When the Tango M2 encounters these situations, it returns a 
reading of 0/0. Microphone placement is critical for reliable operation of the 
Tango M2; there are many places to find cuff placement help. 
a.  See the eLibrary within your Tango M2 for quick cuff placement tutorials. 

You can find the eLibrary under the Main Menu; select Monitor Setup  E-

Summary of Contents for FX-8400

Page 1: ...rement and allows the Tango M2 BP reading to be transferred to the FX 8400 display and reports o SunTech Part Number 91 0010 01 or FUKUDA DENSHI Part Number CJS 03DA5 o Connection Ends FX 8400 Connection Tango M2 Connection Mini DIN 8pin male D sub 9 pin female l ECG Trigger Cable provides the ECG signal from the FX 8400 to the Tango M2 o SunTech Part Number 91 0018 02 or FUKUDA DENSHI Part Number...

Page 2: ...8400 3 Connect the ECG Trigger Cable From the BNC External ECG connection on the rear panel of the Tango M2 To the 1 8th phono plug connection on the side of the FX 8400 RS 232C Connection Part 91 0010 01 or CJS 03DA5 CJS 03DA5 ECG Triger Connection Part 91 0018 02 or CJS 02JB5 RS 232C Connection Part SunTech PN 91 0010 01 Or FUKUDA DENSHI PN CJS 03DA5 ...

Page 3: ...ECG Triger Connection PN 91 0018 02 or FUKUDA DENSHI PN CJS 02JB5 ...

Page 4: ...the Setting button c Touch the External Device button d Select BP Monitor Type to Tango e Touch the To exam button f When a confirmation message appears touch YES FAQs for Tango M2 Q The Tango M2 displays a Status Message What does it mean and what do I do A See the eLibrary within your Tango M2 for quick troubleshooting tips You can find the eLibrary under the Main Menu select Monitor Setup E Lib...

Page 5: ...brightness of the Tango M2 display A You can adjust the contrast of the Tango M2 display by following these steps 1 When the operating screen is displayed press the SELECT button once This will bring up the Main Menu screen 2 Using the UP or DOWN arrows highlight Monitor Setup and press the SELECT button 3 Using the UP or DOWN arrows highlight Brightness and press the SELECT button 4 Using the UP ...

Page 6: ... maintain the accuracy of the Tango M2 s BP readings You will need to contact your local SunTech Medical Service Center for help In addition the following items will be needed to verifycalibration Equipment Required 1 Calibrated electronic manometer or equivalent 2 500mL volume or the Orbit KAdult Plus cuff wrapped around something that will not break or crush no glass 3 Hand Inflation Bulb with b...

Page 7: ...urope the Middle East Africa Asia and the Pacific SunTech Medical Ltd Oakfield Industrial Estate Eynsham Oxfordshire OX29 4TS UK England Phone 44 0 1865 884 234 Fax 44 0 1865 884 235 Email CustomerSupport SunTechMed com ...