FUKUDA DENSHI FX-8400 

Tango M2 Interface Notes 

 

To setup Tango M2 with the FX-8400, simply follow the directions below. 

 

1.  Verify Correct RS-232 and ECG Trigger Cables 

l

 

RS-232 Cable

 – used to communicate with the FX-8400. This cable enables the FX-8400 to prompt 

Tango M2 when it needs a BP measurement, and allows the Tango M2 BP reading to be transferred 

to the FX-8400 display and reports. 

o SunTech Part Number = 91-0010-01 or (FUKUDA DENSHI Part Number CJS-03DA5) 

o Connection Ends: 

FX-8400 Connection 

Tango M2 Connection 

Mini DIN 8pin male 

D-sub 9 pin female 

 

 

 

l

 

ECG Trigger Cable

 – provides the ECG signal from the FX-8400 to the Tango M2. 

o SunTech Part Number = 91-0018-02 or (FUKUDA DENSHI Part Number CJS-02JB5) 

o Connection Ends: 

FX-8400 Connection 

Tango M2 Connection 

1/8” phono plug 

BNC male 

 

 

 

Summary of Contents for FX-8400

Page 1: ...rement and allows the Tango M2 BP reading to be transferred to the FX 8400 display and reports o SunTech Part Number 91 0010 01 or FUKUDA DENSHI Part Number CJS 03DA5 o Connection Ends FX 8400 Connection Tango M2 Connection Mini DIN 8pin male D sub 9 pin female l ECG Trigger Cable provides the ECG signal from the FX 8400 to the Tango M2 o SunTech Part Number 91 0018 02 or FUKUDA DENSHI Part Number...

Page 2: ...8400 3 Connect the ECG Trigger Cable From the BNC External ECG connection on the rear panel of the Tango M2 To the 1 8th phono plug connection on the side of the FX 8400 RS 232C Connection Part 91 0010 01 or CJS 03DA5 CJS 03DA5 ECG Triger Connection Part 91 0018 02 or CJS 02JB5 RS 232C Connection Part SunTech PN 91 0010 01 Or FUKUDA DENSHI PN CJS 03DA5 ...

Page 3: ...ECG Triger Connection PN 91 0018 02 or FUKUDA DENSHI PN CJS 02JB5 ...

Page 4: ...the Setting button c Touch the External Device button d Select BP Monitor Type to Tango e Touch the To exam button f When a confirmation message appears touch YES FAQs for Tango M2 Q The Tango M2 displays a Status Message What does it mean and what do I do A See the eLibrary within your Tango M2 for quick troubleshooting tips You can find the eLibrary under the Main Menu select Monitor Setup E Lib...

Page 5: ...brightness of the Tango M2 display A You can adjust the contrast of the Tango M2 display by following these steps 1 When the operating screen is displayed press the SELECT button once This will bring up the Main Menu screen 2 Using the UP or DOWN arrows highlight Monitor Setup and press the SELECT button 3 Using the UP or DOWN arrows highlight Brightness and press the SELECT button 4 Using the UP ...

Page 6: ... maintain the accuracy of the Tango M2 s BP readings You will need to contact your local SunTech Medical Service Center for help In addition the following items will be needed to verifycalibration Equipment Required 1 Calibrated electronic manometer or equivalent 2 500mL volume or the Orbit KAdult Plus cuff wrapped around something that will not break or crush no glass 3 Hand Inflation Bulb with b...

Page 7: ...urope the Middle East Africa Asia and the Pacific SunTech Medical Ltd Oakfield Industrial Estate Eynsham Oxfordshire OX29 4TS UK England Phone 44 0 1865 884 234 Fax 44 0 1865 884 235 Email CustomerSupport SunTechMed com ...

Reviews: