"Part 2 of 2: High-level disinfecting of transducer for semi-critical
Part 2 of 2: High-level disinfecting of transducer for semi-critical contact
WARNING:
Wear the appropriate personal protective equipment (PPE)
recommended by the chemical manufacturer, such as eyewear and gloves.
High-level disinfectants can cause harm to the patient if not completely
removed from the transducer. Follow the manufacturer’s rinse instructions
to remove chemical residue.
Caution:
Do not skip any steps or abbreviate the disinfecting process in any way.
Do not use unapproved disinfecting agents.
Do not soak the transducer longer than recommended by the chemical
manufacturer.
Do not immerse the transducer connector in any disinfectant solution.
Use only FUJIFILM SonoSite recommended disinfectants. Using a non-
recommended disinfecting solution or incorrect solution strength can damage or
discolor the transducer and void the transducer warranty.
Transducer
Disinfectant
Temperature
Disinfectant
soak duration
Clean water
soak time
UHF22
UHF48
UHF70
Cidex
25°C (77°F)
45 minutes
3 x 1 minute
Compatible high-level disinfectants for semi-critical contact
1 Ensure
you have cleaned the transducer as seen in
system and transducer, disinfecting of system for semi-critical contact"
Step A: Semi-critical contact cleaning and disinfecting of system and
transducer
323
Summary of Contents for VisualSonics Vevo MD
Page 1: ......
Page 2: ......
Page 12: ...12 ...
Page 69: ...System settings 69 ...
Page 70: ...70 System settings ...
Page 77: ...3 Tap DICOM Setup Connectivity 77 ...
Page 146: ...2 Tap User Management in the list on the left 146 System settings ...
Page 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Page 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Page 462: ...462 Index ...
Page 463: ...51370 01 1 0 51370 01 ...