4.5  Endoscope Withdrawal

(1)  When a procedure is over, discharge any excessive air 

from the body cavity.

(2)  Unlock the up-down and left-right knobs.

(3) 

If the balloon is attached, deflate it completely.

[Note]

If the balloon does not deflate while depressing the suction 

valve fully, remove clogs in the balloon channel using the 
cleaning brush (WB2221FW2).

 Reprocessing Manual

 “4.9.4 Brushing the Balloon Channel”

(4)  Straighten the bending section by operating the angulation 

knobs.

Chapter 4  Method of Use

81

EG580UR_E2-50_202B1262871D.indb   81

2016/07/04   16:07:07

Summary of Contents for EG-580UR

Page 1: ...doscope and cautions to be observed when operating it Please read this manual thoroughly before actually operating the ultrasonic endoscope After reading this manual store it nearby the ultrasonic end...

Page 2: ...202B1262871D 2 EG580UR_E2 50_202B1262871D indb 2 2016 07 04 16 06 45...

Page 3: ...precautions and the cleaning disinfection and storage methods for the ultrasonic endoscope If you are a first time user of this product be sure to read this manual before actual operation Also after...

Page 4: ...nctions and damages resulting from negligence of the precautions and operating methods contained in this manual 7 FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from u...

Page 5: ...doscopes EG 580UT and EG 580UR Reprocessing Manual as Reprocessing Manual How to Read This Manual Conventions Used in This Manual This manual uses the following conventions for easier understanding Ge...

Page 6: ...tem configuration Chapter 3 Preparation and Inspection Before Use This chapter describes the inspection and preparation methods to be performed for using the ultrasonic endoscope Chapter 4 Method of U...

Page 7: ...ossible Combinations for Use 26 1 6 1 Accessories 26 1 6 2 Compatible Processor Light Source and Ultrasonic Processor 26 1 6 3 Peripheral Devices 26 1 7 Cautions Warnings 27 1 7 1 Abnormalities during...

Page 8: ...e Ultrasonic Endoscope Processor Light Source Water Tank and Suction Unit 54 3 6 Inspecting Ultrasonic Endoscope 56 3 6 1 Inspecting the Insertion Portion 56 3 6 2 Inspecting the Bending Mechanism 57...

Page 9: ...Troubleshooting 85 5 1 Troubleshooting 85 Main Specification 93 Service 100 Disposal of Electric and Electronic Equipment 101 Index 102 Service Centers 103 9 Contents EG580UR_E2 50_202B1262871D indb...

Page 10: ...performing observation diagnosis and endoscopic treatment of small intestine refer to general contraindications to upper gastrointestinal tract endoscopy and colonoscopy In addition if the general co...

Page 11: ...against electric shock Type BF applied part 3 Degree of explosion protection Use is prohibited in an oxygen rich environment or in a flammable gas atmosphere 4 Degree of waterproof IEC 60529 IPX7 with...

Page 12: ...zation to protect your eye and skin and to prevent infection When using this product for a patient with Creutzfeldt Jakob disease especially variant Creutzfeldt Jakob disease use it exclusively for a...

Page 13: ...is a risk of infection During ultrasonic procedures in which a balloon is not used do not fully depress the suction valve and do not attempt to aspirate patient material into through the balloon evacu...

Page 14: ...infection Use a cleaned and disinfected or sterilized air water valve and suction valve An inadequately cleaned and disinfected or sterilized valve may pose an infection risk Use a cleaned and disinf...

Page 15: ...how to operate the electrosurgical instrument Connect the electrosurgical instruments and electrosurgical generators in accordance with each operation manual Incorrect connection may cause electric sh...

Page 16: ...oves when using an electrosurgical instrument or accessory If not worn there is a risk of thermal injury or electric shock Always keep pacemaker users away from electrosurgical instruments The operati...

Page 17: ...osurgical instruments Doing so may cause thermal injury Operate the instruments within specified output range as per the device s operating instructions Leakage current may cause thermal injury Perfor...

Page 18: ...lation of the balloon Do not use a balloon on patients allergic to latex There is a risk of an anaphylactic reaction Do not supply an excessive amount of air or gas during the procedure as doing so co...

Page 19: ...ended period of time Use the ultrasonic endoscope with a minimum necessary amount of brightness and time while maintaining an appropriate distance When the shutter speed is set to HIGH take care not t...

Page 20: ...ion pressure between 40 and 53 kPa If the suction pressure is too high the ultrasonic endoscope may adhere to mucous membrane resulting in damage to the mucous membrane Do not look directly into the l...

Page 21: ...vices that can be determined by turning the processor off and on we recommend that you may try to correct the interference by one or more of the following measures Change the orientation or position o...

Page 22: ...elow The relevant safety signs are also described 1 5 1 Location of Labels Control Portion Model Label Instrument Channel Diameter Label LG Connector LG Connector Label Label for the Serial Number Lab...

Page 23: ...202B1262871D US Connector Caution Label 23 Chapter 1 Precautions EG580UR_E2 50_202B1262871D indb 23 2016 07 04 16 06 46...

Page 24: ...f manufacture Manufacturer Authorised representative in the European Community Consult instructions for use Temperature limitation Keep dry Non sterile CE marking Degree of waterproof Type BF applied...

Page 25: ...e for autoclave Note 1 This product shall not be treated as household waste Note 2 This symbol indicates that this specific type of medical device contains dibutyl phthalate DBP But the device has lit...

Page 26: ...cessories refer to each operation manual of accessories 1 6 2 Compatible Processor Light Source and Ultrasonic Processor This product is used in combination with the VP 4450HD processor XL 4450 light...

Page 27: ...e monitor Not doing so may cause damage to both the ultrasonic endoscope and body cavities of the patient Wear personal protective equipment when handling an ultrasonic endoscope to prevent infection...

Page 28: ...re reusable devices subject to routine wear and tear Many factors can contribute to the reuse life of any instrument Besides clinical use routine handling as well as repeated cleaning and disinfection...

Page 29: ...he light source Using the ALC Automatic Light Control mode is recommended Refer to the manual of the light source for the ALC mode When this product projects light at the maximum brightness for 2 3 mi...

Page 30: ...ote Figures in parentheses indicate the number of articles Forceps Valve FOV DV7 10 Manuals Operation Manual 1 Reprocessing Manual 1 Carrying Case 1 Waterproof Cap 1 Air Water Valve AW 602 1 Suction V...

Page 31: ...003FW 1 Ventilation Adapter AD 7 1 Cleaning Adapter CA 608 1 Injection Tube WA 007 Channel Plug WA 008 Air Water Channel Cleaning Adapter CA 609 1 US Waterproof Cap WA 7000 1 Chapter 2 Product Overvie...

Page 32: ...treatment Recording video images Printing still images Ultrasonic Endoscope EG 580UR Data Keyboard DK 4450E Processor VP 4450HD Cart Water Tank WT 4 WT 2 Light Source XL 4450 Ultrasonic Processor SU 1...

Page 33: ...DIANCE24 NDS Electrosurgical Unit Note 1 ICC 200 ERBE Note 1 For details refer to the manual of the electrosurgical instrument Keyboard CP 1 Foot Switch FS1 Monitor CDL1909A CDL1576A RADIANCE24 NDS Pr...

Page 34: ...to the ultrasonic processor LG Flexible Portion Contains the light guide air water tube suction tube and cables LG Connector Connects to the light source Air Supply Tube Supplies air from the pump of...

Page 35: ...lid It performs the function of preventing leak or backflow of air Instrument Channel Inlet Inlet for passing forceps or endotherapy device through the instrument channel The forceps valve is attache...

Page 36: ...with the air water valve operation Transducer This portion transmits and receives ultrasonic waves Balloon Attachment Grooves This portion is used to attach the balloon ring Imaging Section Inside Li...

Page 37: ...prevents leak or backflow of air when an endoscopic accessory such as an endotherapy device is used Lid This portion functions as a valve for preventing the leak or backflow of air How to Use the For...

Page 38: ...d to this connector to feed air or fluid to the air water channel of the ultrasonic endoscope Tube for Instrument Suction Channel Connects to the suction channel Air Supply Tube Port Connects to the a...

Page 39: ...the ultrasonic endoscope or to the instrument channel inlet cleaning adapter Instrument Channel Inlet Cleaning Adapter Connects to the instrument channel inlet of the ultrasonic endoscope with the ins...

Page 40: ...failure The bending section is operated by using up down left right angulation knobs Up down Angulation Knob Turn the up down angulation knob in the direction of U to angulate the bending section upw...

Page 41: ...ion knob in the direction of L to angulate the bending section to the left Turn it in the direction of R to angulate the bending section to the right R L R L Chapter 2 Product Overview 41 202B1262871D...

Page 42: ...operated either before or after operating the bending section Note When operating the bending section while the angle is locked angulation knob rotation becomes heavy Up down Angulation Lock Move the...

Page 43: ...Free unlock Allows external force to angulate the bending section freely Rotate the knob in the direction opposite to F to lock the bending section Note Lock Maintains the angle of the bending sectio...

Page 44: ...is pressed halfway the water is supplied When the air water valve is fully pressed water is supplied to the balloon Stopping the hole with a finger Half press Full press Suction valve Suctions water...

Page 45: ...ch For details see the operation manual of the processor Scope Switch 4 Scope Switch 5 Scope Switch 3 Scope Switch 2 Scope Switch 1 For details see the operation manual of the processor Chapter 2 Prod...

Page 46: ...e of the ultrasonic image Note The ultrasonic image is a radial image extending from the point approximately 10 mm ahead of the distal end of the ultrasonic endoscope For details refer to the operatio...

Page 47: ...an create a risk of infection to patients and or end users Be sure to inspect the forceps valve before attachment If any abnormality is found during inspection do not use the product It could be a sou...

Page 48: ...he cap a and hole in the main body b of the forceps valve are free from abnormalities such as tears cracks discoloration etc 2 Attach the lid to the main body of the forceps valve 3 1 3 Attaching the...

Page 49: ...infected or sterilized air water valve and suction valve 3 2 1 Inspecting the Air Water Valve and Suction Valve Visually check that the valves are free from abnormalities such as foreign substances ad...

Page 50: ...s air water valve cylinder and push in the valve firmly Note The air water valve and the air water valve cylinder have a blue mark 2 Attach the suction valve to the ultrasonic endoscope s suction valv...

Page 51: ...eparing Peripheral Devices For details on preparation and inspection of peripheral devices refer to their manuals Chapter 3 Preparation and Inspection Before Use 51 EG580UR_E2 50_202B1262871D indb 51...

Page 52: ...a risk of infection 3 4 1 Preparing the System 1 Move the cart with the processor light source ultrasonic processor and peripherals to the place where ultrasonic endoscope is to be used Note Refer to...

Page 53: ...t daily using sterile water Note When using the balloon the sterile water in the water tank must be deaerated Note Use a cleaned and disinfected or sterilized water tank 5 As a precaution in case of u...

Page 54: ...is no dirt on the light incident distal face and no cracks on the glass of the light incident distal face 2 Insert the LG connector of the ultrasonic endoscope into the ultrasonic endoscope socket on...

Page 55: ...between the suction unit and suction connector of the ultrasonic endoscope 7 Turn on the suction unit and then set the suction pressure of suction source between 40 and 53 kPa 8 Check that the US conn...

Page 56: ...ith both hands to make a semicircle with a diameter of approximately 200 mm referring to the figure at right Then move the apex of semicircle sliding insertion tube for full length Check that the tube...

Page 57: ...turns smoothly When the angulation knob is released the bending section should move toward its neutral position 3 Turn the up down and left right angulation locks in the direction opposite to F and t...

Page 58: ...t may splash 4 With the distal end of the ultrasonic endoscope held in the air press the air water valve completely Check that water is discharged from the balloon water port Note Mind the direction i...

Page 59: ...nd check that the water is sucked in from the balloon water port and suction stops when the finger is released from the valve Note If little or no sterile water is aspirated refer to Chapter 5 Trouble...

Page 60: ...into the light coming from the light guide at the distal end of the ultrasonic endoscope Turn off the lamp before inspecting the objective lens Viewing the light from the light guide directly may dama...

Page 61: ...and check that it is not heated 4 Visually check the transducer the US connector and the cables for abnormalities such as cracks or dents etc Chapter 3 Preparation and Inspection Before Use 61 EG580U...

Page 62: ...ee from cracks and crushes and is securely fixed The side surface of the distal end is free from scratches peeling and abnormal bulging Note If the distal adhesives are missing peeling deteriorated or...

Page 63: ...wiping does not remove cloudiness from objective lens it is possible that ultrasonic endoscope is not sufficiently watertight Perform air tightness test Reprocessing Manual 4 6 Air Tightness Test 8 W...

Page 64: ...ndoscope into water 2 Press the FREEZE key on the keyboard of the ultrasonic processor to cancel the freeze mode 3 Check that an ultrasonic image appears 4 Check that the ultrasonic image is displayed...

Page 65: ...e is a risk of an anaphylactic reaction CAUTION Do not feed more than 5 ml of water into the balloon There is a risk of rupturing the balloon Note It takes approximately 3 seconds to feed in 5 ml of w...

Page 66: ...le water 4 Hold the balloon ring with your fingers and put it around the balloon attachment tool 5 Insert the distal end of the ultrasonic endoscope into the balloon attachment tool until the balloon...

Page 67: ...the amount of water to be fed before use Note Check that there is no water leak from the balloon 8 Press the suction valve fully to suck out all sterile water inside of the balloon Note If water cann...

Page 68: ...Press the suction valve fully and check that all of the sterile water is sucked out and the balloon shrinks Chapter 3 Preparation and Inspection Before Use 68 EG580UR_E2 50_202B1262871D indb 68 2016...

Page 69: ...o could cause an embolism Do not overinsufflate during any clinical procedure to minimize the potential for pneumatic perforation Note Always observe the patient closely If the patient has symptoms su...

Page 70: ...dotherapy devices etc to be used and a spare forceps valve as well Note Use a cleaned and disinfected or sterilized mouthpiece only 4 1 2 Pretreatment of Patient Prepare the patient in the normal endo...

Page 71: ...nic endoscope into the oral cavity Otherwise it may cause pain to the patient Do not forcibly advance or withdraw the ultrasonic endoscope into from the patient It may cause damage to the body lumen b...

Page 72: ...asonic endoscope may be coagulated by the energy of illumination For details on how to use the light save function refer to the operation manual of light source Note If any steam like smoke appears in...

Page 73: ...a patient to lie on examining table in a proper position according to endoscopy procedures 2 Unlock the bending section by turning the up down and left right angulation locks in the direction of F unt...

Page 74: ...he like directly to the insertion portion 5 Press the Lamp button on the light source to turn on the lamp Note If the light source is off press the power button to turn it on 6 Insert the distal end o...

Page 75: ...w the ultrasonic endoscope forcibly Note In case that the bending section does not return or cannot be pulled out easily because it is inverted inside the narrow lumen do not pull it out forcibly 9 In...

Page 76: ...r complete clog of the narrow balloon evacuation channel especially if the ultrasonic endoscope is not cleaned immediately after each procedure and if not reprocessed as per FUJIFILM recommendations 4...

Page 77: ...Note It takes approximately 3 seconds to feed in 5 ml of water Check the amount of water to be fed before use Note Check the endoscopic and ultrasonic images while injecting discharging water into fro...

Page 78: ...nd Note If the balloon does not deflate while depressing the suction valve fully remove clogs in the balloon channel using the cleaning brush WB2221FW2 Reprocessing Manual 4 9 4 Brushing the Balloon C...

Page 79: ...ully when encountering difficulty during insertion There is a risk of damaging the ultrasonic endoscope Note The forceps may not pass through the bending section smoothly In such a case slightly unben...

Page 80: ...field of view 4 Slowly move the forceps to the biopsy site 5 Perform a biopsy by controlling the forceps and the angulation knobs 6 Pull out the forceps slowly and take out the biopsy speci men Chapt...

Page 81: ...ched deflate it completely Note If the balloon does not deflate while depressing the suction valve fully remove clogs in the balloon channel using the cleaning brush WB2221FW2 Reprocessing Manual 4 9...

Page 82: ...02B1262871D 5 Slowly pull out the ultrasonic endoscope 6 Press the Lamp button on the light source to turn off the lamp Chapter 4 Method of Use 82 EG580UR_E2 50_202B1262871D indb 82 2016 07 04 16 07 0...

Page 83: ...2 Roll the balloon slowly toward the distal end of the ultra sonic endoscope to remove Note Do not pinch the transducer with your fingers Note If the balloon is too slippery to be held cover the ballo...

Page 84: ...after use of the ultrasonic endoscope The ultrasonic endoscope is removed after pre cleaning Note For details on pre cleaning and removal of the ultrasonic endoscope refer to the Reprocessing Manual...

Page 85: ...e bending section to unlock release the angulation knobs and then withdraw the ultrasonic endoscope slowly from the patient 2 Reconnect the ultrasonic endoscope 3 5 Connecting Ultrasonic Endoscope 3 S...

Page 86: ...aighten the bending section to unlock release the angulation knobs and then withdraw the ultrasonic endoscope slowly from the patient 2 Reconnect the ultrasonic endoscope 3 5 Connecting Ultrasonic End...

Page 87: ...charge 3 The video signal cable has shorted out or broken 1 2 3 If the problem occurred during treatment stop treatment and remove the endotherapy device from the ultrasonic endoscope Then reset Note...

Page 88: ...the angulation knobs and then withdraw the ultrasonic endoscope slowly from the patient No air and or water delivery 1 Pump is switched off 2 Abnormality in the air water button 3 The air water butto...

Page 89: ...on valve cannot be removed The suction valve or the control portion of the ultrasonic endoscope has been damaged Contact your local FUJIFILM dealer The volume of water supplied to or discharged from t...

Page 90: ...he grip and pull out the endotherapy device from the ultrasonic endoscope 3 Reduce the angle of the bending section slightly and then pull out the endotherapy device from the ultrasonic endoscope 4 Wi...

Page 91: ...ices 2 Turn the power switches ON 3 Increase the amount of water fed to the balloon or adjust the angle of the transducer by operating it so that it comes in close contact with the wall of gastrointes...

Page 92: ...he US connector is poorly connected 1 Stop power supply to the diathermic treatment equipment to restore image output The ultrasonic endoscope is working properly 2 Redo the scope connection 3 5 Conne...

Page 93: ...t Main Unit Endoscopic functions Model EG 580UR Viewing direction 0 Observation range 3 to 100 mm Field of view 140 Distal end diameter 11 4 mm Maximum diameter of insertion portion 12 7 mm Insertion...

Page 94: ...mode Spatial peak temporal average intensity Ispta 720 mW cm 2 assuming the attenuation coefficient of 0 3 dB cm 1 MHz 1 Mechanical Index MI 1 0 or less Frequency 5 MHz 7 5 MHz 10 MHz 12 MHz Operatin...

Page 95: ...any deterioration or abnormality is found in accessories they need to be replaced Accessories cannot be repaired or refurbished Thus if any abnormality is found replace with a new one Consumable Supp...

Page 96: ...tube BF2418SF BF2418FN BF3018FN BF2418LN BF2218A GF2318T GF2418V GF2318B DS2323M DS2323H DC1818R DC2318S WT1720ST Note For combination with accessories other than those above contact your local FUJIFI...

Page 97: ...very low and are not likely to cause any interference in nearby electric equipment Radiated emissions EN 55011 Class A This product is intended for use in medical facilities and commercial facilities...

Page 98: ...wer quality should be that of a typical commercial or hospital Voltage dips short interruptions and voltage variations on power supply input lines EN 61000 4 11 11 5V 218 5V dip For 0 5 cycle 92V 138V...

Page 99: ...with the requirements of EN 60601 1 2 2007 However electromagnetic interference may occur on this product under electromagnetic environment that exceeds its noise level Electromagnetic interference ma...

Page 100: ...s product will be repaired free of charge within warranty guidelines The warranty period for the ultrasonic endoscope excluding accessories is one year after date of purchase Note that the warranty is...

Page 101: ...ectronic equipment By ensuring this product is disposed of correctly you will help prevent potential negative consequences to the environment and human health which could otherwise be caused by inappr...

Page 102: ...ht angulation knob 35 Left right angulation lock 35 LG connector 34 LG connector label 22 Light guide rod 34 M Monitor 33 P Precautions 10 Preparation 47 Printer 33 S S connector 34 Storage environmen...

Page 103: ...in your country USA Fujifilm Medical Systems U S A Inc http www fujifilmendoscopy com 800 385 4666 Australia FUJIFILM Australia Pty Ltd http www fujifilm com au 1800 060 209 Asia FUJIFILM Singapore Pt...

Page 104: ...5 0 DT E2 Printed in Japan FUJIFILM Europe GmbH Heesenstrasse 31 40549 Duesseldorf Germany FUJIFILM Corporation 26 30 Nishiazabu 2 chome Minato ku Tokyo 106 8620 Japan EG580UR_E2 50_202B1262871D indb...

Reviews: