700-SDI-RADAR-300WL-Man Rev 2 11 Apr Mar 2022
30
/48
Part# 21371
X COMMAND
DEFINITION
RESPONSE AND DETAILS
aXGLUN!
Get measurement
unit for level
Response: a+d
In which “d’ is the level measurement unit code:
0 – mm
1 – cm
2 – m
3 – in
4 – ft
d!
Set measurement
unit for level
Replace “d” with desired measurement unit code (see codes
above in Get Measurement Unit Level details)
Response: a+d
aXGDZ0!
Get minimum
dead zone value
Response: a+f
In which “f” =the active zone minimum value.
Sensor will not report a measurement lower than this value.
f!
Set minimum
dead zone value
Replace f with desired minimum dead zone value.
Sensor will not report a measurement lower than this value.
Response: a+f
aXGDZ1!
Get maximum
dead zone value
Response: a+f
In which “f” = the dead zone maximum value.
Sensor will not report a measurement higher than this value.
f!
Set maximum
dead zone value
Replace f with desired dead zone maximum value.
Sensor will not report a measurement higher than this value.
Response: a+f
aXGSHR!
Get sensor height
value
Response: a+f
In which f = sensor height above riverbed
f!
Set sensor height
value
Replace “f” with sensor height above riverbed
Response: a+f
f!
Set current staff
gauge reading
Replace “f” with current staff gauge reading
Device will calculate sensor height above riverbed as:
staff gauge r distance from sensor to water
Response: a+f
aXGLAV!
Get average time
in seconds for
level
Response: a+f
In which “f” = averaging time in seconds
f!
Set average time
in seconds for
level
Replace “f” with desired averaging time in seconds
Response: a+f
aXGDUN!
Get measurement
unit for discharge
Response: a+d
In which “d’ is the code for the discharge measurement unit
0 = m
3
/s
1 = liters
2 = ft
3
/s
d!
Set measurement
unit for discharge
Replace “d” with the code for the desired discharge
measurement unit (see codes above in Get Velocity Discharge
Unit details)