Use the proper power cord with correct attachment plug type. If the power source is 120 V AC,
use a power cord which is a Hospital Grade Power Cord with NEMA 5-15 style plug, labeled for
125 volts AC with UL and C-UL approvals. If the power source is a 240 V AC supply, use the tan-
dem (T blade) type attachment plug with ground conductor power cord that meets the respec-
tive European country’s safety regulations.
A ground post, located on the back of the display, may be used for the purpose of grounding
the display’s chassis. Any such ground must be installed in accordance with applicable electrical
codes. The ground post is shown on the mechanical drawing found in this instructions for use.
Recycling (WEEE Directive 2012/19/EU)
Follow local governing ordinances and recycling plans regarding the recycling or disposal of this
equipment.
Warning:
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the
other equipment should be observed to verify that they are operating normally.
Warning:
Use of accessories, transducers and cables other than those specifi ed or provided
by the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
Warning:
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of this medi-
cal LCD monitor, including cables specifi ed by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Warning:
Using this equipment in the X-ray or magnetic resonance environment could result in
degradation of the performance of this equipment, interference with other equipment or inter-
ference with radio services.
Warning:
The use of cables and/or other accessories with this device, other than those specifi ed,
may result in increased emissions or decreased immunity of this device.
Warning:
This product is not considered physically to connect to HF (High Frequency) electrosur-
gical equipment.
Warning:
Not suitable for use in the presence of a fl ammable anesthetics mixture with oxygen or
with nitrous oxide.