Interfaces

You have the option of using the PC for monitoring and controlling the calibrator via the interfaces (RS232 and USB). In 
control, measurement and automatic mode, a cyclical issue of the current device status can be switched on and off. Ausga-
beintervall ist dabei 1 Sekunde.

 

>

USB

The PC can access through the USB interface on the calibrator (Virtual Com Port). The control via USB is not different 
from the control via RS232.

 

>

RS232

For the connection of RxD, TxD and GND are required. The PC will be connected via a gstraight interface cable (1:1, male à 
female).

 

>

Settings

Push the UP/ DOWN button till SETTINGS is shown on the display and confirm with OK. Then push UP/ DOWN till RS232/ 

USB is displayed. The following settings can be made:

 

>

AKTIV: If a interface is required, choose of the indiviual interface  (USB, RS232 und -------)

 

>

BAUD: Choose of the correct Baudrate (1200, 2400, 4800, 9600, 14400, 19200, 28800, 38400, 56000, 57600)

 

>

DATA-BITS: Default value (8)

 

>

STOP-BITS: Default value (1)

 

>

PARITY-BITS: Default value (N)

 

>

Drivers

You can get the current interface drivers (Windows, Linux, Mac OS) on:

 

>

Homepage URL: http://www.ftdichip.com

 

>

Download URL: http://www.ftdichip.com/Drivers/VCP.htm (FTDI Homepage --> Drivers --> VCPDrivers)

 

>

Installation guide: http://www.ftdichip.com/Documents/InstallGuides.htm (FTDI Homepage --> Documents --> Installati-
on Guides)

 

>

Interface commands

All interface commands are preceded by a colon and the carriage return (CR) complete.Command and parameters must 
be separated by a space. By appending a question mark at the respective command can otherwise be read to changing 
parameters. Receptions commands are acknowledged with „OK“, and missing or incorrect commands with „ERROR“. You 
can get the interfaceprotocol directly from [email protected] anfordern.

Summary of Contents for DPC

Page 1: ...files MENU Customer specific adaption to the respecti ve application Setup Function for different operating modes Operation I O Button to switch on and off the device MENU This button will take you to the setup menu There you can adjust the settings for each mode of operation When you use the CTRL mode the calibrator according to the setup menu to leave the VENT mode LEAK This button performs a LE...

Page 2: ...s the actual value on the LCD Connection Pressure devices have to be connected to the and port of the calibrator Absolute pressure devices have to be connected to the port Setting Press the MENU button Confirm MODE with OK and press UP DOWN until CTRL is shown on the display Confirm with OK button Set the upper limit of the range in the submenu RANGE UP DOWN and 0 100 confirm with OK Set of the un...

Page 3: ...vices to the and port Connect absolute pressure devices to the port Setting Push the MENU button Confirm MODE with OK and push UP DOWN till MESS is displayed Confirm with OK Submenu RANGE and STEPS aren t relevant in the MESS mode Push the UP DOWN buttons to get to the submenu UNIT Choose the relevant unit and confirm with the OK button Push the MENU button to quit Submenu SETTUNGS Push the UP DOW...

Page 4: ...Relevant pressure range UNIT Pressure unit see settings STEPS The pressure range is divided into sections Further settings of AUTO mode can be set in the submenu CONFIGURATION Cycles Set the number of cycles that have to be realized in the AUTO mode T Start Wait for vent of the internal sensor Stop Wait after reaching the maximum pressure T Pause Wait between two cycles AUTO ZERO Automatic zeroing...

Page 5: ...NT is shown left above in the display LEAK leakage In the CTRL mode you it is possible to check connected pressure device for leaks Connection Pressure devices have to be connected to the and port of the calibrator Absolute pressure devices have to be connected to the port Settings To check pressure devices for leaks you have to work in the CTRL mode Enter the desired pressure with the buttons UP ...

Page 6: ...owing settings can be made AKTIV If a interface is required choose of the indiviual interface USB RS232 und BAUD Choose of the correct Baudrate 1200 2400 4800 9600 14400 19200 28800 38400 56000 57600 DATA BITS Default value 8 STOP BITS Default value 1 PARITY BITS Default value N Drivers You can get the current interface drivers Windows Linux Mac OS on Homepage URL http www ftdichip com Download UR...

Reviews: