> AUTO

Store specific profiles in the AUTO mode. Use the AUTO mode to calibrate different pressure devices with idential values. 
The defined profile can be repeated a several times. 

 

>

Connection

Connect differential pressure devices to the + and - port. Connect absolute pressure devices to the + port.

 

>

Settings

 

>

Push the button MENU. Confirm MODUS with OK and push UP/ DOWN till AUTO is displayed. Confirm with OK.

 

>

The submenus RANGE, UNIT and STEPS aren‘t relevant in AUTO mode. 

 

>

Push the buttons UP/ DOWN till SETTINGS is displayed and confirm with OK. Submenu ZERO isn‘t relevant in this 

mode. 

 

>

Push the buttons UP/ Down till AUTO-MODUS is displayed and confirm with OK.

 

>

RANGE: Relevant pressure range

 

>

UNIT: Pressure unit (see settings)

 

>

STEPS: The pressure range is divided into sections.

Further settings of AUTO mode can be set in the submenu CONFIGURATION.

 

>

Cycles: Set the number of cycles that have to be realized in the AUTO mode.

 

>

T Start: Wait for vent of the internal sensor. 

 

>

Stop: Wait after reaching the maximum pressure.

 

>

T Pause: Wait between two cycles

 

>

AUTO-ZERO: Automatic zeroing at the end of every clycle.

Exit of the submenu by pushing the button MENU.

 

>

Use

After exit of the MENU, the calibrator is on wait status. All informations about AUTO mode are shown on the display.  

 

>

Right above: Pressure unit

 

>

In the middle: Phase 1 Sensor is venting (Pressure = 0)

 

>

Left in the middle: First of ten cycles

 

>

Down in the middle: Actual target pressure in per cent

 

>

In the middle above: Target pressure

Push OK button to start the AUTO mode. 

The figure shows the flow chart of the AUTO mode. There you can get detailed informations about the cycles and the 

function.

Summary of Contents for DPC

Page 1: ...files MENU Customer specific adaption to the respecti ve application Setup Function for different operating modes Operation I O Button to switch on and off the device MENU This button will take you to the setup menu There you can adjust the settings for each mode of operation When you use the CTRL mode the calibrator according to the setup menu to leave the VENT mode LEAK This button performs a LE...

Page 2: ...s the actual value on the LCD Connection Pressure devices have to be connected to the and port of the calibrator Absolute pressure devices have to be connected to the port Setting Press the MENU button Confirm MODE with OK and press UP DOWN until CTRL is shown on the display Confirm with OK button Set the upper limit of the range in the submenu RANGE UP DOWN and 0 100 confirm with OK Set of the un...

Page 3: ...vices to the and port Connect absolute pressure devices to the port Setting Push the MENU button Confirm MODE with OK and push UP DOWN till MESS is displayed Confirm with OK Submenu RANGE and STEPS aren t relevant in the MESS mode Push the UP DOWN buttons to get to the submenu UNIT Choose the relevant unit and confirm with the OK button Push the MENU button to quit Submenu SETTUNGS Push the UP DOW...

Page 4: ...Relevant pressure range UNIT Pressure unit see settings STEPS The pressure range is divided into sections Further settings of AUTO mode can be set in the submenu CONFIGURATION Cycles Set the number of cycles that have to be realized in the AUTO mode T Start Wait for vent of the internal sensor Stop Wait after reaching the maximum pressure T Pause Wait between two cycles AUTO ZERO Automatic zeroing...

Page 5: ...NT is shown left above in the display LEAK leakage In the CTRL mode you it is possible to check connected pressure device for leaks Connection Pressure devices have to be connected to the and port of the calibrator Absolute pressure devices have to be connected to the port Settings To check pressure devices for leaks you have to work in the CTRL mode Enter the desired pressure with the buttons UP ...

Page 6: ...owing settings can be made AKTIV If a interface is required choose of the indiviual interface USB RS232 und BAUD Choose of the correct Baudrate 1200 2400 4800 9600 14400 19200 28800 38400 56000 57600 DATA BITS Default value 8 STOP BITS Default value 1 PARITY BITS Default value N Drivers You can get the current interface drivers Windows Linux Mac OS on Homepage URL http www ftdichip com Download UR...

Reviews: