General settings

Press the MENU button and than press the UP / DOWN until you are on the desired sub menu. With OK you go into the 
appropriate sub menu. The following items in the submenu are available: 

 

>

RANGE: Setting of the measurement range (100% value)

 

>

UNIT: Selection of the pressure unit (Pa, hPa, kPa, mbar, bar, Torr, mmHG, inHG, psi, mmH20, inH20)

 

>

STEPS: Setting of the step size in percent.

 

>

MODE: Selection of the operating mode

 

>

CTRL (control): Calibration of pressure sensors and pressure switches

 

>

MESS (measure): Measurement of differential and absolute pressure

 

>

AUTO (automatic): Deposit the pressure profiles

 

>

LANGUAGE: Select language

 

>

SETTINGS: see below

Submenu settings

 

>

ZEROING: On and off the automatic zeroing function (exit the operating menü). Time frequency of the automati   
cal zeroing. 

 

>

RS232/ USB: Select interface

 

>

DISPLAY: Brightness setting

 

>

AUTO: Setting of the range, unit and steps in which the range is divided. 

 

 

>

STEPS UP/ STEPS DOWN: Various values can be assigned. 

 

>

INFO: Informations about the calibrator

Operating Modes And Operation
> CTRL (control)

The CTRL mode is used for calibration of pressure sensors, switches and other pressure equipment. The pressure pump is 
active and regulate the corresponding target pressure. The internal sensor displays the actual value on the LCD. 

 

>

Connection

Pressure devices have to be connected to the + and - port of the calibrator. 
Absolute pressure devices have to be connected to the + port.

 

>

Setting

 

>

Press the MENU button. Confirm MODE with OK and press UP/ DOWN until CTRL is shown on the display. Confirm with 

OK button

 

>

Set the upper limit of the range in the submenu RANGE (UP/ DOWN and 0%/ 100%), confirm with OK

 

>

Set of the unit in the submenu UNIT (UP/ DOWN), confirm with OK

 

>

Set the step sequence (in per cent) in the submenu STEPS (UP/ DOWN and 0%/ 100%) and confirm with OK

 

>

MENU button to exit the submenu

 

>

Use

The unit is located at exit from the menu in VENT mode, ie the sensor of the calibrator is vented. On the middle of the top, 
the target pressure is displayed. On the buttom of the display, the per cent value is shown. When leaving the menu, this 
value is 0%. Switch between target pressure and per cent value with the OK button. 

 

>

The percentage value can be changed with the UP / DOWN (in the fixed step sizes) and using the 0% / 100%.

 

>

The target pressure can only be changed with the UP / DOWN. With 0% / 100% to the appropriate place, the target 
pressure be jumped.

 

>

Example

Target pressure: By pressing the OK button, you can switch between the percentage and target pressure. value. Press 
the OK button until the value (top center display) outlined in bold. Press the 0% or 100% button to jump to the selected 
location. Choose the required value with the UP and DOWN button. Push the OK button to exit target pressure. The marker 
moves to the percentage value. (middle of the display).

Summary of Contents for DPC

Page 1: ...files MENU Customer specific adaption to the respecti ve application Setup Function for different operating modes Operation I O Button to switch on and off the device MENU This button will take you to the setup menu There you can adjust the settings for each mode of operation When you use the CTRL mode the calibrator according to the setup menu to leave the VENT mode LEAK This button performs a LE...

Page 2: ...s the actual value on the LCD Connection Pressure devices have to be connected to the and port of the calibrator Absolute pressure devices have to be connected to the port Setting Press the MENU button Confirm MODE with OK and press UP DOWN until CTRL is shown on the display Confirm with OK button Set the upper limit of the range in the submenu RANGE UP DOWN and 0 100 confirm with OK Set of the un...

Page 3: ...vices to the and port Connect absolute pressure devices to the port Setting Push the MENU button Confirm MODE with OK and push UP DOWN till MESS is displayed Confirm with OK Submenu RANGE and STEPS aren t relevant in the MESS mode Push the UP DOWN buttons to get to the submenu UNIT Choose the relevant unit and confirm with the OK button Push the MENU button to quit Submenu SETTUNGS Push the UP DOW...

Page 4: ...Relevant pressure range UNIT Pressure unit see settings STEPS The pressure range is divided into sections Further settings of AUTO mode can be set in the submenu CONFIGURATION Cycles Set the number of cycles that have to be realized in the AUTO mode T Start Wait for vent of the internal sensor Stop Wait after reaching the maximum pressure T Pause Wait between two cycles AUTO ZERO Automatic zeroing...

Page 5: ...NT is shown left above in the display LEAK leakage In the CTRL mode you it is possible to check connected pressure device for leaks Connection Pressure devices have to be connected to the and port of the calibrator Absolute pressure devices have to be connected to the port Settings To check pressure devices for leaks you have to work in the CTRL mode Enter the desired pressure with the buttons UP ...

Page 6: ...owing settings can be made AKTIV If a interface is required choose of the indiviual interface USB RS232 und BAUD Choose of the correct Baudrate 1200 2400 4800 9600 14400 19200 28800 38400 56000 57600 DATA BITS Default value 8 STOP BITS Default value 1 PARITY BITS Default value N Drivers You can get the current interface drivers Windows Linux Mac OS on Homepage URL http www ftdichip com Download UR...

Reviews: