User Manual

Pressure Calibrator | Type DPC

Notes

 

>

To avoid damage, please read carefully before using

 

>

The maximum allowable pressure shall not be exceeded: DPC 1/ DPC 10/ DPC 100 fivefold overload; DPC 1000  twofold 

overload

 

>

The calibrator is able to overstep and fall below the target pressure about -/+10%. 

 

>

The calibrator may be opened only by or in consultation with FSM Elektronik

Operating Modes

The device has four different modes of operation:

Operating mode

Application

Function

CTRL (control)

Calibration of sensors and other devices.

Pressure is built up and maintained. Target 
pressure may be percentage or absolute. 

MESS (measure)

Measurement of differential and relative 
pressure.

The calibrator analyzes the pressures that 
are connected to the deivces.

AUTO (automatic)

Stored pressure profiles are running.

Pumps builds up the pressure, based on the 

stored pressure profiles.

MENU

Customer specific adaption to the respecti

-

ve application.

Setup-Function for different operating 
modes.

Operation

 

>

I/O: Button to switch on and off the device

 

>

MENU: This button will take you to the setup menu. There you can adjust the settings for each mode of operation. When 
you use the CTRL mode, the calibrator according to the setup menu to leave the VENT mode. 

 

>

LEAK: This button performs a LEAK test of the connected pressure devices.

 

>

ZERO: Button for manual zeroing of the integrated sensor.

 

>

UP: In CTRL mode you can set the absolute or percentage values.; In MENU mode to go into the various submenus. 

 

>

DOWN: In CTRL mode you can set the absolute or percentage values.; In MENU mode to go into the various submenus.

 

>

OK: In CTRL mode you can change between the absolute and percentage value with the OK button. In MENU mode, the 

button serves to confirm each setting. Start the AUTO mode with the OK button. 

 

>

%: In CTRL mode you can use the % button to change the value from 0 to 100. In submenu for setting the decimal.

 

>

The display, you can always check the battery status and the energy (see below).

Allgemeine Einstellungen

1) Allgemeine Einstellung

27

.01.2

016

27

.01.2

016

Summary of Contents for DPC

Page 1: ...files MENU Customer specific adaption to the respecti ve application Setup Function for different operating modes Operation I O Button to switch on and off the device MENU This button will take you to the setup menu There you can adjust the settings for each mode of operation When you use the CTRL mode the calibrator according to the setup menu to leave the VENT mode LEAK This button performs a LE...

Page 2: ...s the actual value on the LCD Connection Pressure devices have to be connected to the and port of the calibrator Absolute pressure devices have to be connected to the port Setting Press the MENU button Confirm MODE with OK and press UP DOWN until CTRL is shown on the display Confirm with OK button Set the upper limit of the range in the submenu RANGE UP DOWN and 0 100 confirm with OK Set of the un...

Page 3: ...vices to the and port Connect absolute pressure devices to the port Setting Push the MENU button Confirm MODE with OK and push UP DOWN till MESS is displayed Confirm with OK Submenu RANGE and STEPS aren t relevant in the MESS mode Push the UP DOWN buttons to get to the submenu UNIT Choose the relevant unit and confirm with the OK button Push the MENU button to quit Submenu SETTUNGS Push the UP DOW...

Page 4: ...Relevant pressure range UNIT Pressure unit see settings STEPS The pressure range is divided into sections Further settings of AUTO mode can be set in the submenu CONFIGURATION Cycles Set the number of cycles that have to be realized in the AUTO mode T Start Wait for vent of the internal sensor Stop Wait after reaching the maximum pressure T Pause Wait between two cycles AUTO ZERO Automatic zeroing...

Page 5: ...NT is shown left above in the display LEAK leakage In the CTRL mode you it is possible to check connected pressure device for leaks Connection Pressure devices have to be connected to the and port of the calibrator Absolute pressure devices have to be connected to the port Settings To check pressure devices for leaks you have to work in the CTRL mode Enter the desired pressure with the buttons UP ...

Page 6: ...owing settings can be made AKTIV If a interface is required choose of the indiviual interface USB RS232 und BAUD Choose of the correct Baudrate 1200 2400 4800 9600 14400 19200 28800 38400 56000 57600 DATA BITS Default value 8 STOP BITS Default value 1 PARITY BITS Default value N Drivers You can get the current interface drivers Windows Linux Mac OS on Homepage URL http www ftdichip com Download UR...

Reviews: