User Manual
Pressure Calibrator | Type DPC
Notes
>
To avoid damage, please read carefully before using
>
The maximum allowable pressure shall not be exceeded: DPC 1/ DPC 10/ DPC 100 fivefold overload; DPC 1000 twofold
overload
>
The calibrator is able to overstep and fall below the target pressure about -/+10%.
>
The calibrator may be opened only by or in consultation with FSM Elektronik
Operating Modes
The device has four different modes of operation:
Operating mode
Application
Function
CTRL (control)
Calibration of sensors and other devices.
Pressure is built up and maintained. Target
pressure may be percentage or absolute.
MESS (measure)
Measurement of differential and relative
pressure.
The calibrator analyzes the pressures that
are connected to the deivces.
AUTO (automatic)
Stored pressure profiles are running.
Pumps builds up the pressure, based on the
stored pressure profiles.
MENU
Customer specific adaption to the respecti
-
ve application.
Setup-Function for different operating
modes.
Operation
>
I/O: Button to switch on and off the device
>
MENU: This button will take you to the setup menu. There you can adjust the settings for each mode of operation. When
you use the CTRL mode, the calibrator according to the setup menu to leave the VENT mode.
>
LEAK: This button performs a LEAK test of the connected pressure devices.
>
ZERO: Button for manual zeroing of the integrated sensor.
>
UP: In CTRL mode you can set the absolute or percentage values.; In MENU mode to go into the various submenus.
>
DOWN: In CTRL mode you can set the absolute or percentage values.; In MENU mode to go into the various submenus.
>
OK: In CTRL mode you can change between the absolute and percentage value with the OK button. In MENU mode, the
button serves to confirm each setting. Start the AUTO mode with the OK button.
>
%: In CTRL mode you can use the % button to change the value from 0 to 100. In submenu for setting the decimal.
>
The display, you can always check the battery status and the energy (see below).
Allgemeine Einstellungen
1) Allgemeine Einstellung
27
.01.2
016
27
.01.2
016