The clamping unit must be operated with the defined disengagement
torque of the control element (cordless screwdriver). This disengage-
ment torque must be set by the clamping system manufacturer. Thus
other cordless screwdrivers, which were not set by Fritsch, must not be
used. If cordless screwdrivers with too high a torque are used, it will
cause irreparable damage to the clamping system for which no liability
can be assumed.
The use of cordless screwdrivers that merely have a lower torque will
result in the sieve stack being clamped with an insufficient clamping
force.
CAUTION!
Follow the separate operating instructions of the cordless
screwdriver manufacturer. These are included with the
device.
When clamping, make sure to lift the TorqueMaster slightly
(illustration 1 + 2). If the toothed belt is not slightly ten-
sioned, it will be crushed in the TorqueMaster (illustration 3
+ 4).
Accessories