6.4 Micro-precision sieving with the ANALYSETTE 3 PRO
6.4.1 Sieving parameters
Parameters
1 micro sieve
Max. 4 micro sieves
Sieving time
20...30 min
60 min
Amplitude in "MICRO"
operating mode
0.1...0.2 mm
0.5 mm
6.4.2 Feed quantity
Depending on particle size and sieve aperture size, it is possible to sieve a
maximum of 200 mg (up to 1 g for coarse sieves of 50 - 100 µm). Deter-
mine the suitable feed quantity through experiment.
NOTICE!
The device must be in "MICRO" operating mode. The
"MICRO" key must light up, otherwise there is a risk of
destroying the sieves.
In the "MICRO" operating mode the amplitude is limited, so that only
values between 0.1 and 0.5 mm can be selected.
NOTICE!
For sieving with micro-precision sieves, the toothed
belts for dry or wet sieving should be removed from the
belt clamping device. (See
clamping the sieves“ on page 23).
6.4.3 Fitting the micro-precision sieves
NOTICE!
Clamp a maximum of 4 micro-precision sieves!
Using the device
- 29 -