Installation
Fresenius Medical Care
T688 Series
4
V1 Aug 2020
At all times, when positioning the chair, care should be taken to ensure that no part of the chair
comes into contact with any equipment or structures, particularly during emergency
procedures.
It is important to ensure that any staff members who are operating the chair have been trained by
the manufacturer or its agent prior to use.
The installation procedure is as follows:
1.
Position the chair. Do not position the chair so that it is difficult to disconnect the chair from
the mains power.
2.
Remove any temporary ties or packaging.
3.
Connect the power cord to an approved power supply for the relevant country
as indicated by
the label.
WARNING
: To avoid risk of an electric shock, this chair must only be connected to supply
mains with protective earth.
4.
Test the chair by taking it through its complete range of movements.
The power cord should be connected to the power point and the power point should be switched
on. The backup battery (if fitted) will recharge automatically and slow to trickle charge when full.
It is important to charge the backup battery for a minimum of 24 hours before use of the chair.