Service
Fresenius Medical Care
T688 Series
13
V1 Aug 2020
Power Supply Cable removal / replacement
Figure 8. Top Frame
•
To remove the power supply cable:
1.
Apply the chair brakes.
2.
Raise the chair to its highest position, if applicable.
3.
Disconnect the mains power supply cable from the wall outlet.
4.
Remove the seat cushion (
Refer to Seat Cushion removal / replacement section on page 9
).
5.
Undo the bolt (
J
) that holds the earth point wire in place (
Figure 8
).
6.
Simultaneoulsly, press-in the red retainer clips on each side of the power cable (
I
), to enable
the removal of the cable (
Figure 8)
.
•
To replace the power supply cable:
1.
Connect the new power cable through the top frame, onto the control box.
2.
Reattach the earth point wire.
3.
Refit the seat cushion.
J
I