FreeStyle Service Manual MN126-1
E
5-23
9.
Remove hour meter as per Section 5.12.
10. Remove cannula nozzle tubing by cutting tie-wrap (use small-size cutter).
Carefully pull tubing from nozzle barb. (Figure 5.32)
Figure 5.32
11.
Disconnect fan connector using needle-nose pliers, carefully pulling upward
on fan connector housing at location J6 on main circuit board.
Figure 5.33
12. Remove six 4-20 self-tapping plastic screws.
(Figure 5.33)