5-16
FreeStyle Service Manual MN126-1
E
Figure 5.23
5.
Remove the plugs from the top and bottom of the new adsorbent beds.
6. Place a 1.5 mm CS X 6 mm ID o-ring on the bottom cap of the adsorbent
bed cap.
7.
Place a 1.5 mm CS X 3.5 mm ID o-ring on the top cap of the adsorbent
beds.
8. Align product manifold with the outlet on the new adsorbent beds and
gently push down until the o-rings are completely seated.
(Figure 5.25)
Figure 5.24
9.
Place center bolt through product manifold and tighten to 15 in-lbs.