Geiger-Müller Counter   513610 

 
 

 

2017-05-05 / HS 

Product manual

  AC513610

   p.

 5/8

 

Dansk 

manual: 

Se p. 1

  

 

Frederiksen Scientific A/S 

Tel. +45 7524 4966 

[email protected] 

Viaduktvej 35 · DK-6870 Ølgod 

Fax +45 7524 6282 

www.frederiksen.eu 

 

 

 

Overview 

 

5 fixed counting periods (Time = 1, 10, 60, 100 

or 300 seconds) - plus manual start and stop 

 

Optional Repeat Mode – repeat 

measurements without interruption. The main 

display shows every completed measurement 

result, while the current count is displayed in 

an auxiliary display 

 

GM tube voltage can be adjusted between 

400 and 700 V, while still reading the counts 

 

GM Sound gives a click from a loudspeaker for 

each pulse recorded 

 

Ready Beep signals end of measurement 

period - especially handy in repeat mode 

 

Very flexible regarding the GM tubes used 

 

May be controlled from a PC. Digital pulse 

output. (Special cables required - option) 

Quick Guide 

Connect

:  The  counter  accepts  tubes  with  BNC 

connectors  or  with  ¼”  Jack.  Connect  only 

one

 

tube. 

Power  On

:  Press   

    to  turn  on  and  off. 

Right  after the counter  is switched  on  it  is  ready 
for  single  measurements  with  a  measurement 
time of 10 s. 

Navigation:

  Find  the  parameter  you  want  to  set 

using the buttons  

  and  

 . The parameter is 

marked with a flashing arrow in the display: 

Time, 

Repeat, GM Voltage, GM Sound, Ready Beep

Setting

: Press  

  to select the current parameter. 

Then set the parameter with  

  and  

 . 

Accept

: Press  

  to accept the new value. 

Measurement

: Start a measurement by pressing  

 .  This also resets the display. 

The measurement stops when the selected time 
expires. If an infinite period is chosen – or if you 
just want to stop prematurely – press  

  again 

to stop. 

Summary of Contents for 513610

Page 1: ...den praktisk i repeat mode Fleksibel med hensyn til det anvendte r r Kan styries fra PC Mulighed for digitale pulser ud Specialkabler ekstraudstyr Quick guide Tilslut T lleren accepterer enten r r med...

Page 2: ...il en igangv rende m ling forts tte efter tryk p indtil den nye m leperiode udl ber Dette vil kikse hvis der v lges en tid som allerede er udl bet Hvis m leresultatet p denne m de g res ugyldigt nulst...

Page 3: ...tet best r af et bogstav evt efterfulgt af et tal De mulige responser fra apparatet afh nger af kommandoen C L s Firmaoplysninger Returner strengen c copyright 2010 A S S Frederiksen D L s registerv r...

Page 4: ...andoen alene returneres den aktuelle h jttaler status Bruges kommandoen med et tal vil h jttaleren tilsluttes afbrydes i h t tallet U0 GM sound fra Ready Beep fra U1 GM sound til Ready Beep fra U2 GM...

Page 5: ...ially handy in repeat mode Very flexible regarding the GM tubes used May be controlled from a PC Digital pulse output Special cables required option Quick Guide Connect The counter accepts tubes with...

Page 6: ...new period expires This will fail when you choose a time that has already expired If the measurement in this way is invalid the display will be reset When a measurement with fixed time is under way t...

Page 7: ...tion protocol Commands to the unit consist of one letter eventually followed by a number The possible responses from the device depend on the command C Read Company Information Returns the string c Co...

Page 8: ...ead If the register is empty when this command is received a value of 1 is returned To ensure that all values are transferred to the PC this register should be read on intervals not longer than the co...